检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:郭治安[1] 赵景婵[1] 党高潮[1] 常建华[1]
出 处:《色谱》2003年第5期524-526,共3页Chinese Journal of Chromatography
摘 要:采用离子抑制色谱法,用反相C18色谱柱,在不经过衍生的情况下,同时测定了过氧乙酸消毒剂中过氧乙酸(PAA)和过氧化氢(H2O2)的含量。该法以酸性磷酸盐为缓冲液,抑制PAA和乙酸的离解,增大了其保留因子,使组分间能得到很好的分离。利用磷酸盐缓冲溶液在210nm以下的紫外吸收强度比水低的特性,降低了基线噪声,提高了过氧化物测定的灵敏度。组分的检出限分别为3 0×10-7mol/L(PAA)和6 2×10-8mol/L(H2O2)(进样量100μL,相当于检出限为2 3ng(PAA)和2 1ng(H2O2))。Peroxyacetic acid (PAA) and hydrogen peroxide (H2O2) in disinfectants were determined simultaneously without derivatization by ion suppression chromatography. A Hypersil ODS column (250 mm×40 mm id, 5 μm) was used with 15 mmol/L phosphate buffer solution (PBS) (pH 22)acetonitrile (82∶18, v/v) as mobile phase. The flow rate was 08 mL/min, and the detection wavelength was 205 nm. In this method, PBS was used to control the dissociation of PAA and HAc,and to increase the retention factors. The baseline noise was reduced as PBS having lower absorption at 200-210 nm than water, and consequently the sensitivity was improved. The limits of detections of PAA and H2O2 were 23 ng and 21 ng, corresponding to PAA and H2O2 concentrations of 30×10-7 mol/L and 62×10-8 mol/L respecively when the injecting volume was 100 μL. The relative standard deviations were less than 05%. The analysis time was within 5 min.
关 键 词:离子抑制色谱法 测定 消毒剂 过氧乙酸 过氧化氢
分 类 号:R187[医药卫生—流行病学] O657.7[医药卫生—公共卫生与预防医学]