机构地区:[1]蚌埠医学院附属医院检验科,安徽蚌埠233004
出 处:《蚌埠医学院学报》2004年第1期69-71,共3页Journal of Bengbu Medical College
摘 要:目的 :对CELL DYN 1 70 0型全自动血球计数分析仪的性能进行临床应用评价。方法 :应用临床标本测定该仪器的精密度、准确度、携带污染率、线性范围。结果 :各项目的批内CV值 :白细胞 (WBC) 1 .9% ,红细胞 (RBC) 1 .6 5 % ,血红蛋白 (Hb)1 .35 % ,红细胞比积 (HCT) 2 .0 % ,平均红细胞体积 (MCV) 0 .93% ,血小板计数 (BPC) 3.2 %。批间CV值 :高值WBC 2 .7% ,RBC 2 .2 1 % ,Hb 1 .86 % ,HCT 2 .8% ,MCV 1 .9% ,BPC 6 .0 % ;正常值 :WBC 2 .1 % ,RBC 1 .9% ,Hb 1 .7% ,HCT 2 .3,MCV1 .8% ,BPC 3.6 % ;低值WBC 2 .0 % ,RBC 2 .1 % ,Hb 1 .6 8% ,HCT 2 .4 % ,MCV 1 .92 % ,BPC 6 .8%。线性范围较宽WBC(0 .5~ 90 .0 )× 1 0 9/L ,RBC(1 .0~ 7.0 )× 1 0 1 2 /L ,Hb(2 5 .0~ 2 0 0 .0 ) g/L、BPC(1 0 .0~ 999.0 )× 1 0 9/L ;携带污染率低 :WBC 0 .1 % ,RBC 0 % ,Hb 0 % ,BPC 0 .5 % ;准确度高 ,EDTAK2 抗凝的静脉血仪器法和常规手工法结果比较差异均无显著性 (P >0 .0 5 )。结论 :该仪器结果准确 ,重复性好 ,操作易于掌握 ,检测速度快 ,交叉污染小 ,且易于保养。静脉血、末梢血均能检测 1Objective:To evaluate the function of the CELL DYN1700 automated hematologic analyser in clinical application.Methods:The accuracy,precision,contamination carrying rate and linearity range were measured using clinical samples.Results:The CV with in the group were as follows:white blood cell(WBC) 1.9 %,red blood cell(RBC) 1.65%,hemoglobin(Hb) 1.35%,hematocrit(HCT) 2.0%,mean corpuscular volume(MCV) 0.93% and blood platelet counts(BPC) 3.2 %.The CV between the groups were as follows:in the high blood value group,WBC 2.7%,RBC 2.21%,Hb 1.86 %,HCT 2.8%,MCV 1.9% and BPC 6.0%;in the normal blood value group,WBC 2.1%,RBC 1.9%,Hb 1.7%,HCT 2.3%,MCV1.8% and BPC 3.6%;in the low blood value group,WBC 2.0%,RBC 2.1 %,Hb 1.68%,HCT 2.4%,MCV 1.92% and BPC 6.8%.The lineariyt range was wide:WBC(0.5 to 90.0)×10 9/L,RBC(1.0 to 7.0)×10 12 /L,Hb(25.0 to 200.0)g/L?BPC(10.0 to 999.0)×10 9/L.The contaimination carrying rate was low:WBC 0.1%,RBC 0%,Hb 0%,BPC 0.5 %.The accuracy of this analyzer was high.The results showed no significant difference between the EDTA K 2 anti freezing vessl blood analyser and conventional detection.Conclusions:This analyzer has the advantages of high accuracy,high repeatability,easy operation,fast detection and little cross pollution.It can test 18 pharaments with vessl blood or finger blood.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222