CELL-DYN1700型自动血细胞计数仪临床应用评价  被引量:1

Evaluation of CELL-DYN1700 automated hematologic analyser in clinical application

在线阅读下载全文

作  者:李峰[1] 杜军[1] 张凡[1] 

机构地区:[1]蚌埠医学院附属医院检验科,安徽蚌埠233004

出  处:《蚌埠医学院学报》2004年第1期69-71,共3页Journal of Bengbu Medical College

摘  要:目的 :对CELL DYN 1 70 0型全自动血球计数分析仪的性能进行临床应用评价。方法 :应用临床标本测定该仪器的精密度、准确度、携带污染率、线性范围。结果 :各项目的批内CV值 :白细胞 (WBC) 1 .9% ,红细胞 (RBC) 1 .6 5 % ,血红蛋白 (Hb)1 .35 % ,红细胞比积 (HCT) 2 .0 % ,平均红细胞体积 (MCV) 0 .93% ,血小板计数 (BPC) 3.2 %。批间CV值 :高值WBC 2 .7% ,RBC 2 .2 1 % ,Hb 1 .86 % ,HCT 2 .8% ,MCV 1 .9% ,BPC 6 .0 % ;正常值 :WBC 2 .1 % ,RBC 1 .9% ,Hb 1 .7% ,HCT 2 .3,MCV1 .8% ,BPC 3.6 % ;低值WBC 2 .0 % ,RBC 2 .1 % ,Hb 1 .6 8% ,HCT 2 .4 % ,MCV 1 .92 % ,BPC 6 .8%。线性范围较宽WBC(0 .5~ 90 .0 )× 1 0 9/L ,RBC(1 .0~ 7.0 )× 1 0 1 2 /L ,Hb(2 5 .0~ 2 0 0 .0 ) g/L、BPC(1 0 .0~ 999.0 )× 1 0 9/L ;携带污染率低 :WBC 0 .1 % ,RBC 0 % ,Hb 0 % ,BPC 0 .5 % ;准确度高 ,EDTAK2 抗凝的静脉血仪器法和常规手工法结果比较差异均无显著性 (P >0 .0 5 )。结论 :该仪器结果准确 ,重复性好 ,操作易于掌握 ,检测速度快 ,交叉污染小 ,且易于保养。静脉血、末梢血均能检测 1Objective:To evaluate the function of the CELL DYN1700 automated hematologic analyser in clinical application.Methods:The accuracy,precision,contamination carrying rate and linearity range were measured using clinical samples.Results:The CV with in the group were as follows:white blood cell(WBC) 1.9 %,red blood cell(RBC) 1.65%,hemoglobin(Hb) 1.35%,hematocrit(HCT) 2.0%,mean corpuscular volume(MCV) 0.93% and blood platelet counts(BPC) 3.2 %.The CV between the groups were as follows:in the high blood value group,WBC 2.7%,RBC 2.21%,Hb 1.86 %,HCT 2.8%,MCV 1.9% and BPC 6.0%;in the normal blood value group,WBC 2.1%,RBC 1.9%,Hb 1.7%,HCT 2.3%,MCV1.8% and BPC 3.6%;in the low blood value group,WBC 2.0%,RBC 2.1 %,Hb 1.68%,HCT 2.4%,MCV 1.92% and BPC 6.8%.The lineariyt range was wide:WBC(0.5 to 90.0)×10 9/L,RBC(1.0 to 7.0)×10 12 /L,Hb(25.0 to 200.0)g/L?BPC(10.0 to 999.0)×10 9/L.The contaimination carrying rate was low:WBC 0.1%,RBC 0%,Hb 0%,BPC 0.5 %.The accuracy of this analyzer was high.The results showed no significant difference between the EDTA K 2 anti freezing vessl blood analyser and conventional detection.Conclusions:This analyzer has the advantages of high accuracy,high repeatability,easy operation,fast detection and little cross pollution.It can test 18 pharaments with vessl blood or finger blood.

关 键 词:血细胞计数 自动血球计数仪 

分 类 号:R446.113[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象