红外光谱仪多点定标方法及环境温度校正  被引量:15

CALIBRATION METHODS OF INFRARED SPECTRUM RADIOMETER AT VARIOUS TEMPERATURES AND BACKGROUND TEMPERATURE ADJUSTMENT

在线阅读下载全文

作  者:黄烨[1] 方勇华[1] 荀毓龙[1] 熊伟[1] 乔延利[1] 

机构地区:[1]中国科学院安徽光学精密机械研究所,安徽合肥230031

出  处:《红外与毫米波学报》2004年第2期131-134,共4页Journal of Infrared and Millimeter Waves

摘  要:在红外光谱仪的定标中 ,为了减小误差 ,可以选取多温度点的测量来进行定标 ;此外对于测量较弱红外辐射的光谱仪 ,环境温度对定标结果也有不可忽视的影响 .本文主要讨论利用测量多温度点的标准黑体进行定标的方法。In calibration of a infrared spectrum radiometer, we calibrate an infrared radiometer by measuring on various temperatures of the blackbody in order to reduce error. For the radiometer measuring weak radiation, the background temperature has unavoidable effect on calibration. The methods of calibration by measuring the standard blackbody at different temperature was discussed. The experimentation results were analysed, and the effect of the background temperature on the calibration result was given.

关 键 词:红外光谱仪 多点定标 环境温度校正 定标误差 线性拟合算法 

分 类 号:TN214[电子电信—物理电子学]

 