作 者:季钢[1] 程功[1] 万声贤 王晓莉[1] 刘其胜[1] 张正华[1]
机构地区:[1]合肥市妇婴保健院
出 处:《安徽医学》1993年第6期7-8,共2页Anhui Medical Journal
摘 要:采用Minolta-101经皮胆红素测定仪测定生后1-15天的足月新生儿74例。同时用偶氮法测定其血清总胆红素。经皮测定胆红素(TcB)指数与血清总胆红素二者之间呈显著正相关(r=0.872,P<001)。TcB是一种操作简便,无创伤性测定新生儿胆红素的新技术,可作为新生儿高胆红素血症的筛选方法,并能准确及时地监测胆红素的动态变化。This paper reports the use of a thanscutaneous Bilirubinometer-Minolta 101 for determination of Bilirttbin level in 74 cases full term newborn, aged 1-15 days Simultaneous control with J-G method was performed. Significant positive correlation between transcutaneous bilirubinometry (TCB) index and serum bilirubin concentration was present (r=0.872 P<0.01). TCB is a new simple and noninvasive technique for the meaasurement of Biliru bin level in the newborn.It may be used as a screening method in diagnosis of hyperbilirubinemia in newborn. It is quite accurate and variation in bilirubia levels cart be observed promptly.