Reliability of the K-Force Muscle Controller Dynamometer on Eccentric and Isometric Hip Adduction Strength  

Reliability of the K-Force Muscle Controller Dynamometer on Eccentric and Isometric Hip Adduction Strength

在线阅读下载全文

作  者:Christos Pippas Antonis Emmanouilidis Stefanos Karanasios George Koumantakis George Gioftsos Christos Pippas;Antonis Emmanouilidis;Stefanos Karanasios;George Koumantakis;George Gioftsos(Department of Physiotherapy, University of West Attica, Athens, Greece)

机构地区:[1]Department of Physiotherapy, University of West Attica, Athens, Greece

出  处:《Open Journal of Therapy and Rehabilitation》2024年第3期224-235,共12页康复医学(英文)

摘  要:Hip adduction strength tests are commonly used in clinical practice to provide an accurate diagnosis of groin injuries. Athletes with reduced adductor muscle strength are at risk of developing groin injuries. Our study aimed to evaluate the relative and absolute test-retest reliability of the side-lying eccentric hip adduction strength test and the long-lever adduction squeeze test using the K-Force hand-held dynamometer. Twenty physically active male individuals with a mean age (±SD) of 30.7 (±7.3) years were included. Both tests presented excellent test-retest reliability (Intraclass Correlation Coefficient: 0.77 - 0.95). The best and mean scores of the eccentric and isometric tests presented the smallest test-retest variation (MDC%: 12.8 - 14.9 and MDC%: 14.6 - 18.7, respectively). Our study showed that the K-Force dynamometer has excellent reliability for assessing hip adduction strength in two different testing positions. We suggest the best and mean of three repetitions for clinical practice as they present the lowest variability. Further research evaluating its clinimetric properties in different populations and gender is recommended.Hip adduction strength tests are commonly used in clinical practice to provide an accurate diagnosis of groin injuries. Athletes with reduced adductor muscle strength are at risk of developing groin injuries. Our study aimed to evaluate the relative and absolute test-retest reliability of the side-lying eccentric hip adduction strength test and the long-lever adduction squeeze test using the K-Force hand-held dynamometer. Twenty physically active male individuals with a mean age (±SD) of 30.7 (±7.3) years were included. Both tests presented excellent test-retest reliability (Intraclass Correlation Coefficient: 0.77 - 0.95). The best and mean scores of the eccentric and isometric tests presented the smallest test-retest variation (MDC%: 12.8 - 14.9 and MDC%: 14.6 - 18.7, respectively). Our study showed that the K-Force dynamometer has excellent reliability for assessing hip adduction strength in two different testing positions. We suggest the best and mean of three repetitions for clinical practice as they present the lowest variability. Further research evaluating its clinimetric properties in different populations and gender is recommended.

关 键 词:ADDUCTOR GROIN Strength Testing DYNAMOMETER RELIABILITY 

分 类 号:O17[理学—数学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象