Establishment of Diagnostic Reference Levels for Childhood Head CT in the District of Abidjan in Côte d’Ivoire  

Establishment of Diagnostic Reference Levels for Childhood Head CT in the District of Abidjan in Côte d’Ivoire

在线阅读下载全文

作  者:Koutoua Joseph Anouan Raïssa-Michelle Kabas Jean-Paul Kouao Aboubakar Sidiki Diabaté Koutoua Joseph Anouan;Raïssa-Michelle Kabas;Jean-Paul Kouao;Aboubakar Sidiki Diabaté(Nuclear Physics and Radiation Protection Team, Training and Research Unit, Sciences of the Structures of Matter and Technology, University Flix Houphouet-Boigny of Abidjan, Abidjan, Cte dIvoire;Department of Basic and Bioclinical Sciences, Training and Research Unit, Medical Sciences, University Flix Houphouet-Boigny of Abidjan, Abidjan, Cte dIvoire;Department of Radiodiagnosis and Medical Imaging, University Hospital of Treichville, Abidjan, Cte dIvoire)

机构地区:[1]Nuclear Physics and Radiation Protection Team, Training and Research Unit, Sciences of the Structures of Matter and Technology, University Flix Houphouet-Boigny of Abidjan, Abidjan, Cte dIvoire [2]Department of Basic and Bioclinical Sciences, Training and Research Unit, Medical Sciences, University Flix Houphouet-Boigny of Abidjan, Abidjan, Cte dIvoire [3]Department of Radiodiagnosis and Medical Imaging, University Hospital of Treichville, Abidjan, Cte dIvoire

出  处:《World Journal of Nuclear Science and Technology》2025年第1期45-52,共8页核科学与技术国际期刊(英文)

摘  要:Computed tomography is an indispensable X-ray imaging modality used to diagnose numerous pathologies, but it can also involve the delivery of high ionizing radiation doses harmful to the health of patients. This study aims to survey the level of radiation doses delivered to child patients during head exams in CT imaging to set up the Dosimetric Reference Levels (DRLs), a routine dose optimization tool, based on data acquired at the University Hospital of Angré (UHA), the University Hospital of Treichville (UHT) and the Polyclinic Hospital Farah (Farah) for optimizing procedures in Ivorian hospitals. Prospectively performed on 334 CT images of 186 child patients, this study was carried out on CT systems such as Hitachi Scenaria, Sinovision Insitum, and Philips Incisive used respectively at UHA, UHT and Farah. Children’s scan data were classified into four age bands: vol or dose-length product as DLP) value, whatever the hospital, increases with respect to the age of child patients. Based on the 75th percentile of the whole dose distributions, the DRLs of the CTDIvol is 54.37 mGy whatever the age groups and those of the DLP with respect to age bands are 1224.55 mGy∙cm, 1414.06 mGy∙cm, 1632.24 mGy.cm and 1544.57 mGy∙cm, respectively. The averaged values of CTDIvol and DLP smaller than the corresponding DRLs values suggest that practices in our three facilities are optimized. However, comparing our results with those from different international studies, we see that the CTDIvol and DLP values obtained in the present work are higher. These results suggest additional surveys to ensure our DRLs values and efforts from radiologists, imaging technicians and medical physicists to strengthen clinical procedures for the radiation protection of children undergoing CT scans in Côte d’Ivoire.Computed tomography is an indispensable X-ray imaging modality used to diagnose numerous pathologies, but it can also involve the delivery of high ionizing radiation doses harmful to the health of patients. This study aims to survey the level of radiation doses delivered to child patients during head exams in CT imaging to set up the Dosimetric Reference Levels (DRLs), a routine dose optimization tool, based on data acquired at the University Hospital of Angré (UHA), the University Hospital of Treichville (UHT) and the Polyclinic Hospital Farah (Farah) for optimizing procedures in Ivorian hospitals. Prospectively performed on 334 CT images of 186 child patients, this study was carried out on CT systems such as Hitachi Scenaria, Sinovision Insitum, and Philips Incisive used respectively at UHA, UHT and Farah. Children’s scan data were classified into four age bands: vol or dose-length product as DLP) value, whatever the hospital, increases with respect to the age of child patients. Based on the 75th percentile of the whole dose distributions, the DRLs of the CTDIvol is 54.37 mGy whatever the age groups and those of the DLP with respect to age bands are 1224.55 mGy∙cm, 1414.06 mGy∙cm, 1632.24 mGy.cm and 1544.57 mGy∙cm, respectively. The averaged values of CTDIvol and DLP smaller than the corresponding DRLs values suggest that practices in our three facilities are optimized. However, comparing our results with those from different international studies, we see that the CTDIvol and DLP values obtained in the present work are higher. These results suggest additional surveys to ensure our DRLs values and efforts from radiologists, imaging technicians and medical physicists to strengthen clinical procedures for the radiation protection of children undergoing CT scans in Côte d’Ivoire.

关 键 词:Computed Tomography Childhood Head Imaging Radiation Protection Diagnostic Reference Levels 

分 类 号:R81[医药卫生—放射医学]

 