Non-cycloplegic spherical equivalent refraction in adults:comparison of the double-pass system,retinoscopy,subjective refraction and a table-mounted autorefractor  被引量:3

在线阅读下载全文

作  者:Meritxell Vilaseca Montserrat Arjona Jaume Pujol Elvira Peris Vanessa Martìnez 

机构地区:[1]Centre for Sensors,Instruments and Systems Development(CD6),Technical University of Catalonia [2]University Vision Centre (CUV),Department of Optics and Optometry,Technical University of Catalonia

出  处:《International Journal of Ophthalmology(English edition)》2013年第5期618-625,共8页国际眼科杂志(英文版)

基  金:Spanish Ministry of Education and Science(No.DPI2008-06455-C02-01);European Union and the Spanish Agency for International Cooperation(AECI)(No.D/030286/10)

摘  要:AIM:To evaluate the accuracy of spherical equivalent(SE) estimates of a double-pass system and to compare it with retinoscopy,subjective refraction and a table mounted autorefractor.METHODS:Non-cycloplegic refraction was performed on 125 eyes of 65 healthy adults(age 23.5±3.0 years) from October 2010 to January 2011 using retinoscopy,subjective refraction,autorefraction(Auto kerato refractometer TOPCON KR-8100,Japan) and a double pass system(Optical Quality Analysis System,OQAS,Visiometrics S.L.,Spain).Nine consecutive measurements with the double-pass system were performed on a subgroup of 22 eyes to assess repeatability.To evaluate the trueness of the OQAS instrument,the SE laboratory bias between the double pass system and the other techniques was calculated.RESULTS:The SE mean coefficient of repeatability obtained was 0.22D.Significant correlations could be established between the OQAS and the SE obtained with retinoscopy(r=0.956,P【0.001),subjective refraction(r=0.955,P【0.001) and autorefraction(r=0.957,P【0.001).The differences in SE between the double-pass system and the other techniques were significant(P【0.001),but lacked clinical relevance except for retinoscopy;Retinoscopy gave more hyperopic values than the double-pass system-0.51±0.50D as well as the subjective refraction-0.23±0.50D;More myopic values were achieved by means of autorefraction 0.24±0.49D. CONCLUSION:The double-pass system provides accurate and reliable estimates of the SE that can be used for clinical studies.This technique can determine the correct focus position to assess the ocular optical quality.However,it has a relatively small measuring range in comparison with autorefractors(-8.00 to +5.00D),and requires prior information on the refractive state of the patient.AIM:To evaluate the accuracy of spherical equivalent(SE) estimates of a double-pass system and to compare it with retinoscopy,subjective refraction and a table mounted autorefractor.METHODS:Non-cycloplegic refraction was performed on 125 eyes of 65 healthy adults(age 23.5±3.0 years) from October 2010 to January 2011 using retinoscopy,subjective refraction,autorefraction(Auto kerato refractometer TOPCON KR-8100,Japan) and a double pass system(Optical Quality Analysis System,OQAS,Visiometrics S.L.,Spain).Nine consecutive measurements with the double-pass system were performed on a subgroup of 22 eyes to assess repeatability.To evaluate the trueness of the OQAS instrument,the SE laboratory bias between the double pass system and the other techniques was calculated.RESULTS:The SE mean coefficient of repeatability obtained was 0.22D.Significant correlations could be established between the OQAS and the SE obtained with retinoscopy(r=0.956,P<0.001),subjective refraction(r=0.955,P<0.001) and autorefraction(r=0.957,P<0.001).The differences in SE between the double-pass system and the other techniques were significant(P<0.001),but lacked clinical relevance except for retinoscopy;Retinoscopy gave more hyperopic values than the double-pass system-0.51±0.50D as well as the subjective refraction-0.23±0.50D;More myopic values were achieved by means of autorefraction 0.24±0.49D. CONCLUSION:The double-pass system provides accurate and reliable estimates of the SE that can be used for clinical studies.This technique can determine the correct focus position to assess the ocular optical quality.However,it has a relatively small measuring range in comparison with autorefractors(-8.00 to +5.00D),and requires prior information on the refractive state of the patient.

关 键 词:double-pass system optical quality RETINOSCOPY autorefraction subjective refraction accuracy repeatability TRUENESS 