PDMS基质微型管萃取-热解析-GC联用测定水中有机物  被引量:3

Determination of trace organic compound in water by a PDMS tube extraction coupled with thermal desorption/GC

在线阅读下载全文

作  者:王丽丽[1] 丁梨慧[1] 陈聪[1] 潘再法[1] 莫卫民[1] 

机构地区:[1]浙江工业大学化学工程与材料学院,浙江杭州310032

出  处:《浙江工业大学学报》2005年第5期551-554,共4页Journal of Zhejiang University of Technology

基  金:浙江省自然科学基金资助项目(Y404079)

摘  要:选用聚二甲基硅氧烷(PDMS)基质微型管为固相微萃取的萃取介质,研究与考察了其与热解析-GC联用对水中痕量有机物的萃取解析特性与快速分析方法.在最佳萃取解析条件下,PDMS基质微型管对5 mL水样中痕量萘、联苯和菲的萃取性能良好,热解析完全.方法的线性范围为30~1000μg/L,线性度高(r>0.992),最低检测限<3 μg/L.雨水样品中萘、联苯和菲(100 μg/L)的加标回收率为70%~85%,RSD(n=3)为3%~14%.结果表明:该方法具有萃取效率高,简单快速、基质影响小,费用低廉等优点.The extraction and desorption performance of a selected polydimethylsiloxane tube coupled with thermal desorption/G-C was studied by analysing trace organic compound in water samples. Under the optimum extraction and desorption conditions, the wide linear concentrations (30 -1000 μg/L) and good correlation coefficients(r〉0. 992)were achieved and the low limits of detection was bellow 3 μg/L for naphthalene, biphenyl and phenanthrene in distilled water. The recoveries of naphthalene, biphenyl and phenanthrene added in rain water (100 μg/L) were between 70 and 85%, and RSDs (n=3) were between 3 and 14 %. The results showed that the polydimethylsiloxane tube extraction coupled with thermal desorption-GC has the advantage of high extraction efficiency, good baseline, low analytical expense as well as simple and fast procedure.

关 键 词:聚二甲基硅氧烷基质微型管 固相微萃取 水样品痕量分析 气相色谱 

分 类 号:O657.71[理学—分析化学]

 