在线血液透析滤过液中细菌内毒素定量法研究  被引量:6

Kinetic turbidimetric limulus assay of bacterial endotoxin in the replacement solution of on-line hemodiafiltration

在线阅读下载全文

作  者:劳海燕[1] 林秋晓[1] 杨敏[1] 梁馨苓[1] 史伟[1] 李慧[2] 冯聚锦 

机构地区:[1]广东省人民医院药学部,广东广州510080 [2]广东药学院,广东广州510240 [3]湛江安度斯生物有限公司,广东湛江524000

出  处:《南方医科大学学报》2006年第2期242-244,共3页Journal of Southern Medical University

基  金:广东省科技计划项目(2003C33704)~~

摘  要:目的应用动态比浊法鲎实验定量测定血液透析滤过液中的细菌内毒素含量。方法对血液透析滤过液中定量添加标准内毒素进行干扰预试验,将血液透析滤过液原液及稀释成5、10、20倍液,定量添加标准内毒素,其回收率分别为 58.17%、106.7%、99.00%和98.79%;从中筛选出最佳的稀释倍数为10倍,进行3个批号血液透析滤过液(040408、 040511、040527)正式干扰试验,血液透析滤过液中定量添加标准内毒素,其回收率分别为76.32%、99.00%和96.24%。结果标准内毒素使用1.00、0.125、0.0156 Eu/ml,将血液透析滤过液稀释成10倍,滤过液中定量添加标准内毒素,回收率均为50%~200%,认为血液透析滤过液对试验无干扰作用,可用于样品检查。结论动态比浊法可以测定血液透析滤过液的细菌内毒素含量。Objective To determine bacterial endotoxin in the replacement solution of on-line hemodiafiltration (on-line HDF) using kinetic turbidimetric limulus test. Methods and Results Validation test was performed with the replacement solution of on-line HDF in which quantified standard endotoxin was added. The recovery rates ofendotoxin from the replacement solution and its dilutions at 1/5, 1/10, and 1/20 were 58.17%, 106.7%, 99.00% and 98.79%, respectively, suggesting that the optimal dilution was at 1/10. Standard endotoxin was added into the replacement solution of on-line HDF of 3 batches (04040g, 040511,040527), and the recovery rates in their dilution at 1/10 were 76.32%, 99.00% and 96.24%, respectively. The standard endotoxin in the working curve was 1.00, 0.125, and 0.0156 Eu/ml (endotoxin unit/ml), and the dilution at 1/10 of the replacement solution is effective to eliminate the interference in limulus test. Conclusion Kinetic turbidimetric limulus test provide a means to detect endotoxin in the replacement solution of on-line HDF.

关 键 词:动态比浊法鲎试验 血液透析滤过液 细菌内毒素 干扰试验 

分 类 号:R927.11[医药卫生—药学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象