摘  要:目的:寻找穿透性小梁切除术后患者眼压测量的简单方法。方法:随机选取行穿透性小梁切除术后3d的患者120例120眼。分别用Goldmann眼压计和Tono-pen眼压计测量眼压,比较两种眼压测量值的差异。结果:120眼中,Goldmann眼压计所测量眼压的平均数值14.69±4.60mmHg。Tono-pen所测眼压的平均数值14.72±4.62mmHg。经统计学检验,P>0.05,Tono-pen眼压计与Goldman眼压计的测量结果无显著性差异。对于穿透性小梁切除术后的眼压测量,Tonopen眼压计与Goldman眼压计的测量结果有高度相关性。Tono-pen眼压计可以准确地估计眼压值。结论:对于穿透性小梁切除术后的的患者眼压在7~32mmHg间时Tono-pen眼压计可以作为一种准确测量眼压的简单方法。AIM: To determine whether Tono-pen tonometer is a good and simple method to measure patients who had undergone trabeculectomy by comparing the intraocular pressure (IOP) values obtained by Goldmann tonometer and Tono-pen tonometer. METHODS: A total of 120 patients underwent trabeculectomy were selected, their IOP was measured at 3 day after surgery with Goldmann tonometer and Tono-pen tonometer. RESULTS: The mean IOP values obtained by Goldmann tonometer and Tono-pen tonometer were 14.69± 4.60mmHg and 14.72± 4.62 mmHg, respectively, with no statistically significant difference (P〉0.05). There was high correlation between the reading of the two instruments from 7mmHcj to 32mmHg.CONCLUSION: Tono-pen tonometer can be used as a good and simple methodon patients undergone trabeculectomy when their IOP is 7-32mmHg.

关 键 词:穿透性小梁切除术 眼压 Tono—pen眼压计 GOLDMANN眼压计 

分 类 号:R779.6[医药卫生—眼科]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象