根管长度电测仪用于乳牙根管长度测量的体外实验  

In vitro measurement accuracy of an electronic apex locator in primary teeth with and without root resorption

在线阅读下载全文

作  者:曹筝[1] 陈莉丽[2] 王莹[3] 

机构地区:[1]浙江大学医学院附属邵逸夫医院牙科,浙江大学邵逸夫临床医学研究所,浙江杭州310016 [2]浙江大学医学院附属第二医院口腔科 [3]浙江大学医学院附属口腔医院

出  处:《全科医学临床与教育》2007年第5期378-379,396,共3页Clinical Education of General Practice

摘  要:目的探讨根管长度电测仪用于根尖有无生理性吸收的乳牙根管长度测量的准确性。方法选取40颗离体乳前牙,根据根尖区有无吸收情况分为两组,分别置于体外模型中,用RootZX根管长度电测仪测量其工作长度,将其与在5倍放大镜下测量出的真正结果相比较。结果在±0.5mm范围内,根尖有吸收组电测长度准确率为85.00%,无吸收组为90.00%。虽然测量值较真实值略低,但两者差异无统计学意义(χ2=0.23,P>0.05)。结论体外测量结果表明,乳牙根尖生理性吸收的存在不会影响电测法的准确性。Objective To assess the accuracy of an electronic device (RootZX) for determining root canal length in primary teeth with and without root resorption. Methods Forty extracted primary incisors (twenty with visible apical root resorption and twenty without) were used in this study. The primary teeth were mounted in an experimental apparatus separately. After achieving ideal access, working length measurements were obtained using the RootZX, comparing the electronic measurements with the direct visual measurements under the magnifying glasses ( × 5). Results Within 0.5mm the RootZX was 85.00% accuracy in group with root resorption and 90.00 % accuracy in group without resorption. No significant difference was found between the two groups (χ^2 = 0.23, P 〉 0.05), and the electronic measurements had the slight underestimation shown in the study. Conclusions The results suggest that the accuracy of RootZX is not adversely affected by the presence of resorption in primary.

关 键 词:根管长度测量仪 乳牙 工作长度 吸收 

分 类 号:R788[医药卫生—口腔医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象