流动注射分析中混合釜体积与进样量的关系  

Relationship between the Volume of Mixing Tanks and Injected Sample Volume in Flow Injection Analysis

在线阅读下载全文

作  者:吴宁生[1] 吴国胜[1] 史文娟[1] 

机构地区:[1]中国科学技术大学化学系,合肥230026

出  处:《分析化学》1997年第7期765-767,共3页Chinese Journal of Analytical Chemistry

摘  要:在流路系统和载流流速一定且进样体积不大于反应器体积的0.45倍时,系统混合釜总体积Vm与进样体积Sv的关系可表示为:Vm=(a+bSv)/(1+Sv),a和b为与实验条件有关的常数。单釜体积为定值,与进样体积无关,在流路系统和进样体积一定时,混合釜总体积为定值,与载流流速无关。For a given manifold and flow rate and the sample volume is not greater than 45/100 reactor volume, the relationship between total volume of mixing tanks V. and inject- ed sample volume Sv. can be expressed as: Vm= (a+bSv)/(1+Sv),where a and b are coeffi- cients which depend on experimental conditions, and the volume of single tank is a constant which is independent of the sample volume. For a given manifold and samp1e volume, the to- tal volume of mixing tanks is a constant which is independent of the carrier flow rate.

关 键 词:流动注射分析 串联釜模型 混合釜 进样量 

分 类 号:O657.3[理学—分析化学]

 