摘 要:研制的多道程控刺激仪能够根据操作者要求,以多种模式向电极阵列送出指定幅度和脉宽的脉冲。刺激仪由单片机定时器控制时间参数,由CPLD控制空间参数,由电源控制幅度参数。摒弃了传统变压器隔离的方法,采用光电耦合以及浮地等设计,以达到生理研究需要的安全标准。According to the instruction of the operator, the designed programmable multichannel stimulator will deliver stimuli with an assigned amplitude and a proper width to the electrode-array in various modes. Each stimulation time is controlled by the internal timer of the MCU, while the stimulated electrodes' spatial parameter is controlled by the CPLD. Having abandoned conventional isolation using the transformer, we have adapted optical coupling and stable float ground connection to achieve the safety standard required by physiological measurements.