AcrySof ReSTOR人工晶状体上镊子夹痕2例(英文)  

Forceps imprint in the AcrySof ReSTOR IOL Optic

在线阅读下载全文

作  者:叶向彧[1] 曹景[2] 陶津华[1] 王于蓝[1] 盛耀华[1] 

机构地区:[1]上海交通大学医学院附属新华医院眼科,中国上海市200092 [2]中国山东省青岛市第八人民医院眼科,266100

出  处:《国际眼科杂志》2008年第7期1315-1318,共4页International Eye Science

摘  要:本文报道了两例在植入时出现镊子夹痕的AcrySof Re-STORIOL患者。病例1进行了IOL置换,电镜扫描结果显示AcrySof ReSTORIOL衍射结构保持良好。病例2术后视力、对比敏感和波前相差检查未见明显异常。我们的结论是AcrySof ReSTORIOL光学区上的镊子夹痕不会导致视力下降。We describe two cases in which a forceps imprint developed in the AcrySof ReSTOR IOL optic while inserting these IOLs into the cartridge with straight clamping forceps. In case 1 ,the AcrySof ReSTOR IOL was explanted and observed under scanning electron microscopy (SEM). The SEM showed that the step design of ReSTOR Multifocal IOL was well maintained. In case 2, visual acuity, contrast sensitivity and wavefront measurements were performed and no specific changes were found. Strong evidence does not exist that suggests the on-axis forceps imprint can significantly compromise visual acuity.

关 键 词:镊子夹痕 人工晶状体 多焦 

分 类 号:R779.6[医药卫生—眼科]

 