G6PD试剂盒在Roche modulor 010自动化分析仪上的应用及评价  被引量:1

Evaluation of the G6PD reagents kit on Roche modulor 010 auto-biochemistry analyzer

在线阅读下载全文

作  者:曾建明[1] 张烁[2] 陈茶[1] 张茂威[1] 庞鑫[1] 马骥[1] 王美兰[1] 庄俊华[1] 

机构地区:[1]广东省中医院检验科,广东广州510120 [2]广东医学院检验学院,广东广州524023

出  处:《现代医药卫生》2008年第16期2387-2389,共3页Journal of Modern Medicine & Health

摘  要:目的:评价G6PD试剂盒在Roche modulor 010自动化分析仪的应用情况。方法:通过重复测定检测系统的精密度,按EP6-A文件稀释并检测线性,同时检测质控物评价其稳定性。另外,检测疑似HbH病人的G6PD活性。结果:MHb-RT高值和低值标本批内、批间CV分别是4%、3%和2%、8%;G6PD/6PGD比值的决定系数R2分别为0.999;质控物CV值为2%。3例疑似HbH病人标本检测结果正常。结论:此试剂盒的精密度和线性较好,能够满足临床常规检验需要。Objective:To evaluate the domestically produced C, tPD kit on Roche modulor 010 auto-biochemistry analyzer.Methods:Selected one high and one low MHb-RT value samples for precision test, the linear relationship was checked according to the protocol of EP6-A document. Quality control was performed everyday.3 highly suspected HbH samples were assayed. Results:The higher value sample GPD/6PGD ratio CV was within-run 4%, between-run 2% ,and the lower value sample CV was within-run 3%,between-run 8% respectively. The determination coefticient of G6PD/6PGD ratio was 0.999. The sequential 10 days CV results of quality control material were 2%.3 highly suspected HbH samples' GPD activity assayed was normal. Conclusion:Precision and linear of C, tPD kit could fit the demand of clinical work.

关 键 词:G6PD 精密度 线性关系 ROCHE modulor 010自动化分析仪 

分 类 号:R446[医药卫生—诊断学]

 