计算机辅助股骨头表面假体植入精确度的实验研究  被引量:4

Accuracy of computer-aided femoral head surface prostheses placement

在线阅读下载全文

作  者:喻忠[1] 王黎明[1] 桂鉴超[1] 蒋纯志[1] 侯明夫[1] 姚京东[1] 徐燕[1] 

机构地区:[1]南京医科大学附属南京第一医院骨科,江苏南京210006

出  处:《医学研究生学报》2008年第9期933-936,I0001,共5页Journal of Medical Postgraduates

摘  要:目的:研究非影像手术导航系统对股骨头表面假体植入精确度的影响。方法:8具尸体(16髋)在导航系统下进行股骨头表面置换术,术中设定髋最佳植入柄颈角为0,°根据导航系统显示的柄干角植入假体。术后根据X线片测定假体植入的角度。对术中、术后获得的假体数据进行统计学分析。结果:术中、术后获得的假体柄颈角、柄干角差异无统计学意义(P>0.05)。结论:非影像手术导航系统下可以精确地植入假体,减少假体松动的发生,具有重要的临床价值。Objective: To study the accuracy of the computer-aided surgical navigation system in femoral head surface component placement. Methods:Femoral head resurfacing operations were performed on l b hips of 8 human cadavers using the non-image based surgical navigation system. The femoral component was placed in a best stem-neck angle of 0° as designed and the prostheses planted according to the stemgraft angle shown by the navigation system. The postoperative component position was measured by X-ray. The data obtained during and after the operation were subjected to statistical analyses. Results: There were no statistically significant differences between the data obtained at two different times (P 〉 0.05). Conclusion :The femoral head surface prostheses can be planted accurately by the non-image based surgical navigation system, which reduces the possibility of loose prostheses and has an important clinical value.

关 键 词:髋关节 表面置换术 手术导航系统 

分 类 号:R687.3[医药卫生—骨科学]

 