眼眶内木质异物CT和MRI诊断  被引量:6

CT and MRI of intraorbital wooden foreign bodies

在线阅读下载全文

作  者:程敬亮[1,2] 岳松伟[1,2] 任翠萍[1,2] 齐云秋 高剑波[1,2] 姜曦[1,2] 施光普[1,2] 

机构地区:[1]河南医科大学第一附属医院放射科 [2]河南医科大学第一附属医院眼科

出  处:《河南医科大学学报》1998年第2期21-24,共4页Journal of Henan Medical University

基  金:河南省科技攻关项目

摘  要:将16枚兔眶内木质异物和3例(4枚)患者眶内木质异物的CT和MRI资料进行比较分析,探讨CT和MRI对眶内木质异物的诊断价值。16枚兔眶内木质异物包括干杨木块5枚、干竹片6枚、青枝柳条5枚,异物大小为1.5mm×1mm×1mm~9mm×4.2mm×4.2mm,埋置异物后2h内完成CT和MRI。3例患者4枚异物中2枚为木块,2枚为竹签,大小为3mm×2mm×1mm~25mm×7mm×7mm。伤后至CT、MRI检查时间为20~60d。结果:16枚兔眶内异物CT检出12枚,CT值介于-145~63.8Hu;MRISE序列T1WI、PDWI和T2WI分别检出11枚、8枚和16枚,Flash2d序列全部检出;3例患者4枚异物CT仅检出1枚,MRI全部检出。结果提示:MRI诊断眶内木质异物优于CT,MRISE序列T2WI和Flash2d序列是检出眶内木质异物的优选序列;お? A preliminary investigation of 16 pieces of intraorbital wooden foreign bodies in rabbits and 4 pieces in 3 eyes traumatic patients was carried out by CT and MRI.The experimental inserted foreign bodies included 5 pieces of dry poplar,6 pieces of bamboo and 5 pieces of fresh willow.Their sizes were from 1.5 mm×1 mm×1 mm to 9 mm×4.2 mm×4.2 mm.CT and MRI were performed within 2 hours after the foreign bodies were implanted.There were 2 pieces of wooden and 2 pieces of bamboo in the 4 injured eyes from 3 patients which ranged from 3 mm×2 mm×1 mm to 25 mm×7 mm×7 mm in size.CT and MRI were carried out 20~60 days after the traumas.The results showed that 12 out of 16 experimental foreign bodies from rabbits were discovered by CT,and CT attenuation value was from -145 to 63.8 Hu. All the intraorbilal foreign bodies of rabbits were picked up by Flash2d sequence or T2WI of SE sequence on MRI,but only 11 and 8 pieces by T1WI and PDWI of SE sequence respectively.Only one out of 4 foreign bodies from 3 patients was discovered by CT,while all of them were picked up by MRI.The findings sugges that MRI is superior to CT in the diagnosis of intraorbital wooden foreign bodies,and Flash2d sequence and T2WI of SE sequence should be the choice sequence.Since long lasting wooden foreign body of the eye may resorb water from its environment,this leads to an increasement of its density on CT and signal intensity on MRI.

关 键 词:眼异物 木质异物 CT 磁共振成像 

分 类 号:R779.140.4[医药卫生—眼科] R816.97[医药卫生—临床医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象