检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]中国第一汽车集团公司技术中心,吉林长春130011
出 处:《冶金分析》2011年第5期71-73,共3页Metallurgical Analysis
摘 要:样品用盐酸、过氧化氢溶解,在少量硫酸存在下,以盐酸和氢溴酸挥锡,电感耦合等离子体原子发射光谱法测定铅和镉的含量。优化了仪器的工作参数,对溶样酸和锡基体等影响因素进行了试验。本法测铅和镉的线性范围分别为0.50~5.00μg/mL和0.050~0.50μg/mL,检出限分别为13.2μg/L和0.9μg/L,样品测定结果的相对标准偏差(n=6)均小于4%,用标准加入法测得回收率分别为97%~108%和90%~96%。The solder samples were dissolved in hydrochloric acid and hydrogen peroxide.Then,the tin was removed by hydrochloric acid-hydrobromide in presence of little sulfuric acid,and the contents of lead and cadmium were determined by inductively coupled plasma atomic emission spectrometry(ICP-AES).The working parameters of ICP instrument were optimized.The influencing factors such as acid for dissolving sample and tin matrix were tested.The linear ranges in this method were obtained between 0.50-5.00 μg/mL(for Pb) and 0.050-0.50 μg/mL(for Cd).The detection limits were 13.2 μg/L for Pb and 0.9 μg/L for Cd,respectively.The relative standard deviations(RSD,n=6) were less than 4 %.The recoveries rate for lead and cadmium obtained by standard addition method were 97 %-108 % and 90 %-96 % respectively.
关 键 词:电感耦合等离子体原子发射光谱法 焊锡 铅 镉
分 类 号:TG115.33[金属学及工艺—物理冶金]
正在载入数据...