分化型甲状腺癌^(131)I全身显像采集条件的选择  

Acquisition Conditions Optimization of ^(131)I Whole Body Scan for Differentiated Thyroid Cancer

在线阅读下载全文

作  者:刘金华[1] 冯会娟[1] 孙云钢[1] 刘伟英[1] 黄柳华[1] 冼嘉朗[1] 欧阳伟[1] 

机构地区:[1]南方医科大学珠江医院核医学科,广州510282

出  处:《医疗卫生装备》2012年第2期73-73,84,共2页Chinese Medical Equipment Journal

摘  要:目的:131I全身显像采集条件的优化选择。方法:选择分化型甲状腺癌患者131I全身显像,用不同的速度采集。结果:采集时间长、速度慢ROI组对比于采集时间短ROI组有统计学意义。结论:选择采集时间为200 s的图像较理想,可以提高仪器的工作效率,图像均匀清晰。To choose the optimized acquisition conditions of 131I-whole body scan. Methods Patients with differentiated thyroid cancer underwent 131I-whole body scan with different speed, For the scan of region of interest (ROI), the difference between the group with slower acquisition speed and the group with faster acquisition speed was significant. The optimized acquisition condition is 200 seconds, using which the instrument efficiency can be improved, and the image will be clear and even.

关 键 词:分化型甲状腺癌 131I全身显像 采集速度 感兴趣区 

分 类 号:R736.1[医药卫生—肿瘤]

 