检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]武汉水利电力大学化学与环境工程系,湖北武汉430072 [2]华南理工大学应用化学系,广东广州510641
出 处:《分析测试学报》2000年第2期67-69,共3页Journal of Instrumental Analysis
摘 要:建立了用荧光分光光度计作为流动注射分析的检测器 ,用自行研制的流通池测定垃圾渗出液中苯胺的新方法。方法的线性范围为0.07~0.5mg/L ,检出限为0.05mg/L。用于垃圾渗出液的测定 ,结果与HPLC法无显著性差异 ,回收率在96 %~104 %之间 ,该法具有选择性好、灵敏度高、重现性好的特点。A new method for aniline determination in rubbish percolate is described. Fluorescence spectrophotometer was used as the FIA detector. A flow pipe developed by ourselves serves as the FIA flow cell. The linear range of this method is 0.07~0.5 mg·L-1, whereas its detection limit is 0.05 mg·L-1. No significant discrepancy between the proposed method and the standard HPLC method was observed, when it was applied to rubbish percolate analysis. The recovery is in the range of 96%~104%. This method has the advantages of good selectivity, high sensitivity and good precision.
分 类 号:X132[环境科学与工程—环境科学] O625.631[理学—有机化学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.3