电容压力传感器与电子血压计的结构原理及电子血压计的校准  被引量:2

Structure principle of capacitive pressure transducer and sphygmomanometer

在线阅读下载全文

作  者:王黎明[1] 苏艳玲[2] 王森[3] 

机构地区:[1]白山市中心医院设备科,吉林白山134300 [2]白山市图书馆,吉林白山134300 [3]白山市中心医院神经内科,吉林白山134300

出  处:《中国医学装备》2013年第3期44-46,共3页China Medical Equipment

摘  要:目的:通过对电容压力传感器与电子血压计的结构、原理进行分析,修复受损的电容压力传感器,并对修复后的血压计进行校准。方法:利用常见的正常血压计做为标准压力计,使两血压计处于同一压力水平来校准血压计。结果:电容压力传感器修复后经过对血压计的校准达到其使用标准。结论:电容压力传感器在电子血压计中相对为易损元件且参数易改变,本文所采用的修复方法及校准电子血压计的方法易于普及。Objective: By analyzing the structure and principle of capacitive pressure transducer and electronic sphygmomanometer, the study solves the problems of repairing damaged capacitive pressure transducer and calibration of sphygmomanometer after repair. Methods: This study uses two ordinary sphygmomanometers as the standard manometer, and by making them the same pressure level to calibrate the sphygmomanometer. Results: By calibrating the sphygmomanometer, the capacitive pressure transducer reaches the using standard after repair. Conclusion: In the electronic sphygmomanometer, capacitive pressure transducer is a relatively vulnerable device. Its parameter is easy to change. Comparing with the standard capacitive pressure transducer, this method is easy to promote and popularize.

关 键 词:电容压力传感器 电子血压计 校准 频率调制 

分 类 号:R197.39[医药卫生—卫生事业管理]

 