检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]广东省潮州市质量计量监督检测所,广东潮州521011 [2]广东省质量监督不锈钢器皿和厨具检验站(潮安),广东潮州521011
出 处:《现代科学仪器》2013年第1期152-154,共3页Modern Scientific Instruments
摘 要:依据JJF1059—1999《测量不确定度评定与表示》,利用真空直读光谱仪测定不锈钢食具容器化学成分含量,以铬元素为例,通过建立数学模型,从人、机、料、法、环、溯六方面分析评定铬元素含量的测量不确定度。分析各不确定度的分量,得出校准曲线和标准样品的均匀性对测量不确定度影响最大的结论,为今后仪器操作者如何提高检测准确性提供理论参考。On the basis of JJF1059-1999 evaluation and expression of uncertainty in measurement,using vacuum direct reading spectrometer for determination of chemical composition content in stainless steel food containers,with Cr element as an example,through the establishment of the mathematical model,the measurement uncertainty of chromium content was evaluated from six factors including man,machine,material,method,central,back analysis.Analyzing the uncertainty components,a calibration curve and standard sample homogeneity of uncertainty in measurement for influential conclusion were drawn.It can provide the theory reference for future equipment operators how to improve the accuracy of testing.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.114