气相色谱-正化学电离质谱法检测婴儿奶嘴中N-亚硝胺物质的迁移量  被引量:15

GC-MS(PCI)Determination of Amount of N-Nitrosamines Migrated from Infant Nipple

在线阅读下载全文

作  者:杨荣静[1] 卫碧文[1] 高欢[1] 于文佳[1] 高鹏 

机构地区:[1]上海出入境检验检疫局,上海200135 [2]岛津企业管理(中国)有限公司,上海200052

出  处:《理化检验(化学分册)》2013年第12期1423-1427,共5页Physical Testing and Chemical Analysis(Part B:Chemical Analysis)

摘  要:提出了气相色谱-正化学电离质谱联用技术检测婴儿奶嘴中12种N-亚硝胺物质的迁移量的方法。样品经人工唾液提取后,过经二氯甲烷活化的活性炭固相萃取柱净化。在气相色谱分离中用DB-Wax色谱柱为固定相,在质谱分析中采用全扫描和选择离子监测模式。12种N-亚硝胺物质的峰面积与质量浓度均在0.01~5.0mg·L^-1范围内呈线性关系,方法的测定下限(10S/N)均为0.001mg·kg^-1。以奶嘴样品为基体做加标回收试验,测得12种N-亚硝胺物质的回收率在89.3%~95.6%之间,相对标准偏差(n=7)在3.6%~7.6%之间。A method of Cd2-MS(PC1) for the determination of amount of 12 N-nitrosamines migrated from infant nipple was proposed. The sample was extracted with artificial saliva solution,and the extract was purified by passing through activated carbon SPE column which was pre-activated by dichloromethane. The DB-Wax column was used for GC separation, and the full-scanning mode and selected ion monitoring mode were adopted in MS determination. Linear relationships between values of peak area and mass concentration of the 12 N-nitrosamines were kept in the same range of 0. 01- 5. 0 mg ·^L^-1, with same lower limit of determination (10S/N) of 0. 001 mg · kg^-1. Using nipple sample as matrix, test for recovery was made by standard addition method, giving recovery of the 12 N nitrosamines in the range of 89. 3%-95. 6%, with RSD's (n=7)in the range of 3. 6%- 7.6%.

关 键 词:气相色谱-正化学电离质谱法 N-亚硝胺 婴儿奶嘴 

分 类 号:O657.7[理学—分析化学]

 