利福霉素钠注射液中环氧丙烷、二甘醇测定方法的研究  

Determination of propylene oxide and diethylene glycol in rifamycin sodium injection by GC

在线阅读下载全文

作  者:姚兰[1] 刘杨[1] 张亚杰[1] 

机构地区:[1]辽宁省食品药品检验所,沈阳110023

出  处:《中国抗生素杂志》2014年第1期49-51,70,共4页Chinese Journal of Antibiotics

摘  要:目的建立利福霉素钠注射液中环氧丙烷与二甘醇的测定方法。方法采用气相色谱法,聚乙二醇为固定液的毛细管柱为色谱柱,程序升温,进样口温度为230℃;检测器为氢火焰离子化检测器(FID),检测器温度为250℃;载气为氮气,柱流量1.1mL/min;分流比为20:1。结果2种杂质在所考察的浓度范围内具有良好的线性,线性范围:环氧丙烷0.00166~0.0332mg/mL,r=0.9982;一缩二乙二醇0.00215-0.0430mg/mL,r=0.9999。平均回收率分别为100.3%、98.2%。定量限分别为环氧丙烷0.376μg/mL,一缩二乙二醇2.12gg/mL。结论方法准确可靠,适用于利福霉素钠注射液中环氧丙烷、一缩二乙二醇杂质的测定。Objective To establish a GC method for the determination of propylene oxide and diethylene glycol in rifamycin sodium injection. Methods The GC was equipped with hydrogen flame-ionazition detector, the detector temperature was 250~C, the inlet temperature was 230℃. Temperature programming and the capillary chromatographic column coated with polyethylene glycol was used in the test, the carrier gas was nitrogen, the column flow was 1. 1 mL/min; and the split ratio was 20:1. Results The 2 kinds of impurities had a good linear relation in the range of 0.00166-0.0332mg/mL, r=0.9982 (propylene oxide) and 0.00215-0.04300mg/mL, r=0.9999 (diethylene glycol). The everage recovery was 100.3% and 98.2% and the quantitative limits were 0.376 and 2.12μg/mL, respectively. Conclusion The method was accurate and reliable, suitable for determination of impurities of propylene oxide and diethylene glycol in rifamycin sodium injection.

关 键 词:二甘醇 环氧丙烷 GC 利福霉素钠注射液 

分 类 号:R927.1[医药卫生—药学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象