出 处:《临床误诊误治》2014年第1期95-97,共3页Clinical Misdiagnosis & Mistherapy
摘 要:目的用SIRIUS 3D角膜地形图眼前节分析系统和A超角膜测厚仪测量中央角膜厚度,并比较其结果差异。方法连续选择拟行准分子激光手术者366例(732只眼),分别应用SIRIUS 3D角膜地形图和TOPCON-3000A超仪测量中央角膜厚度,比较两者测定值及其相关性。结果 A超所测角膜厚度平均(534.02±1.099)μm,SIRIUS3D角膜地形图测量角膜厚度平均(533.20±1.068)μm,二者比较差异有统计学意义(t=2.468,P=0.000),两种测量方法所测结果呈显著正相关(r=0.954,P=0.000)。结论 SIRIUS 3D角膜测厚仪和A超角膜测厚仪各有其优势及劣势,应两种方法结合使用,准确评估角膜厚度,以保证手术安全性。Objective To compare the difference in measurement of central corneal thickness (CCT) by SIRIUS 3D corneal topography of anterior segment analysis system and A-type ultrasonic corneal pachymeter.Methods A total of 366 patients (732 eyes),who would undergo excimer laser surgeries,were continuously selected.The CCT was measured respectively by Sirius 3D corneal topography and TOPCON-3000 A type ultrasonic corneal pachymeter,and the measured value and their correlation were compared.Results The mean CCT values were (534.02 ± 1.099) μm measured by A-type ultrasonic corneal pachymeter,and (533.20 ± 1.068) μm measured by SIRIUS 3D corneal topography,and the differences were statistically significant (t =2.468,P < 0.001) ; 2.The results measured by the two methods showed a high correlation (r =0.954,P =0.000) ; there were positive correlations in results measured by the two methods (r =0.954,P =0.000).Conclusion SIRIUS 3D corneal topography and A-type ultrasonic corneal pachymeter have their own advantages and disadvantages.To evaluate the corneal thickness exactly,the two methods should be combined to guarantee the operative safety.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.179