Cobas C501全自动生化分析仪的检测性能验证  被引量:4

Assessment of the Performance of Cobas C501 Automatic Biochemical Analyzer

在线阅读下载全文

作  者:张丽[1] 蔡鹃 管晓媛[1] 段兵[1] 黄一玲[1] 张波 田蕾[1] 李一石[1] 

机构地区:[1]中国医学科学院北京协和医学院阜外心血管病医院卫生部心血管药物临床研究重点实验室,北京100037 [2]北京洛奇临床检验所有限公司,北京102206

出  处:《标记免疫分析与临床》2014年第2期204-207,共4页Labeled Immunoassays and Clinical Medicine

基  金:国家12.5"重大新药创制"科技重大专项-<心血管创新药物临床研究技术平台建设>(2012ZX09303-008-001);国家临床重点专科建设项目-<卫生部重点实验室项目>

摘  要:目的以总蛋白(TP)为例,考察Roche Cobas C501全自动生化分析仪的精密度、准确度和可报告范围,以验证其检测性能。方法参考美国临床和实验室标准化协会(CLSI)EP5-A2、EP6-A等文件,使用C501的原装配套试剂及校准品,检测总蛋白(TP)的精密度、准确度和可报告范围,用EP Evaluator软件统计分析。结果 TP两水平精密度检测结果的批内标准差(S)、总的S均不超过验证值,精密度可以接受;在对照系统与实验系统之间,20个患者血清样本的检测结果的误差指数(EI)均在±1.00之间,准确度可以接受;覆盖可报告范围的5个水平样本中的4个结果呈线性,线性可接受。结论 Cobas C501分析仪TP项目的精密度、准确度、可报告范围达到了厂家声明的检测性能,此分析仪的检测性能得到了验证。Objective To verify the precision,accuracy and linearity of Cobas C501 automatic biochemical analyzer by using total protein (TP) as a reference.Methods According to EP5-A2,EP6-A etc from Clinical and Laboratory Standards Institute (CLSI),TP was analyzed with Roche original reagent and calibrator to verify Cobas C501 automatic biochemical analyzer precision,accuracy and linearity.The EP evaluator software was applied for statistics analysis.Results In the 2 levels of TP,both of within standard deviation (S) and total S were less than verification value.The precision was acceptable.20 specimens were measured by control instrument and experiment instrument; all of error indexes (EI) were in ± 1.00 range.The accuracy verification passed.In linearity evaluation,the results of four of the five specimens over the reportable range were linear which indicating linearity was acceptable.Conclusion The precision,accuracy and linearity of TP detected by Cobas C501 system reach the performance that the manufacturer claimed.

关 键 词:精密度 准确度 线性 性能评价 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象