质谱测量与分析系统的性能测试  被引量:2

Performance test of gas pressure measuring and analysis system for mass spectrometer

在线阅读下载全文

作  者:曹曾[1] 黄向玫[1] 蔡潇[1] 许正华[1] 田培红 崔成和[1] 高霄雁[1] 

机构地区:[1]核工业西南物理研究院,成都610041

出  处:《核聚变与等离子体物理》2014年第2期135-140,共6页Nuclear Fusion and Plasma Physics

基  金:国家磁约束核聚变能研究专项(2010GB108002)

摘  要:对所研制的质谱测量与分析系统进行了性能测试。烘烤后系统的极限真空达到了7.3×10-7Pa,总漏气率为3.84×10-9Pa.m3.s-1。上下球室之间的气体压力,以及漏孔的漏气率与高分辨四极质谱计的氦分压力均呈现了良好的线性关系。结果满足动态流导法真空校准的技术要求,为氘、氦混合气体的测量与分析提供了条件。The performance test of gas pressure measuring and analysis system for quadrupole mass spectrometer was carried out. The ultimate vacuum is 7.3×10-7Pa after the system was baked, and total leak rate is 3.84×10-9Pa.m3.s-1. The pressure relationship between upper and lower chamber is linear dependence. So is the relationship of the helium leak rate and partial pressure. The results meet the design requirements of dynamic flow conductance, and provide basic conditions for mass spectrograph analysis.

关 键 词:极限真空 漏气率 线性关系 动态流导法 

分 类 号:TL628[核科学技术—核技术及应用]

 