A/B超交替测量眼轴用于人工晶体度数计算  被引量:2

Axial Eye length measurement with alternative A/B scan for IOL power calculation

在线阅读下载全文

作  者:顾真寰[1] 沈慧英[1] 盛敏杰 

机构地区:[1]中国上海同济大学医学院附属铁路局中心医院眼科,200072

出  处:《美中国际眼科杂志》2001年第1期46-48,共3页

摘  要:目的 探索并发性白内障玻璃体混浊时准确测量眼轴的新方法。方法 用A超测量角膜到晶体后囊的距离,利用B超直观性测量晶体后囊到视网膜的距离,两者相加得出眼轴长度用于人工晶体计算,可避免将玻璃体混浊误认为眼后壁。结果 与常规A超测量法相比,术后残余屈光在1D以上的比例减少12%。结论 本测量法设计合理,准确性高,实用性强。Objective To explore new means to measure axial eye length intercurrent cataract and inpatients with vitreous turbidity . Methods The distance between the corneal and posterior capsular. was measured by A scan, and the distance between the posterior capsular and retinal was measured by was B scan measured.The avial eye length was reached by adding these two results and was available for IOL calculation. Results Compared with routine A scan measurement, the cases with postoperative unexpected refractive error of 1 dioptre or above were reduced by 12% .Conclusion The new measurement is reasonable,accurate and practicabe.

关 键 词:超声波测量 眼轴 并发性白内障 

分 类 号:R779.66[医药卫生—眼科]

 