便携式人工心肺辅助装置动物实验研究  被引量:2

Performance of a wearable artificial cardiopulmonary support device in a sheep model

在线阅读下载全文

作  者:刘洋[1] 金振晓[1] 董小超[1] 魏旭峰[1] 杨剑[1] 程亮[1] 俞世强[1] 

机构地区:[1]第四军医大学西京医院心脏外科,西安710032

出  处:《中国体外循环杂志》2015年第2期119-122,共4页Chinese Journal of Extracorporeal Circulation

基  金:军队医学科技青年培育项目(13QNP129)

摘  要:目的初步考察便携式人工心肺辅助装置的在体性能。方法应用肝素涂层中空纤维氧合技术与磁悬浮离心泵装置整合构建小型化、便携式心肺辅助装置。40 kg实验羊(n=6)在全麻下第五肋间左前外侧切口,分别置入右心房插管连接至人工心肺辅助装置流入道,降主动脉插管连接至装置流出道开始转流,4.0 L/min运转12 h,于转流前和转流后每2 h检测装置流入道及流出道血氧饱和度,氧分压,二氧化碳分压等,计算装置氧合效率。取静脉血检测游离血红蛋白检含量评价装置溶血性能并检测血小板激活率。结果 12 h中,装置流出道氧饱和度保持98%以上,氧合效率稳定。4.0 L/min流量运行12 h,游离血红蛋白含量均在20 g/L以下,血小板激活均小于8%。结论应用肝素涂层中空纤维氧合器与磁悬浮离心血泵构建人工心肺辅助装置在体性能可靠。Objective To evaluate the performance of a wearable artificial cardiopulmonary support devicein vivo.Methods Heparin-coating hollow fiber and axial pump were used to setup a wearable artificial cardiopulmonary support device. Six 40-kg sheepwere used in the study. The left anterolateral thoracotomy were performed after general anesthesia. Then the veno-arterial connectionswere setup with cannulation in the right atrium and descending aorta. The saturation, partial oxygen pressure and partial carbon dioxidepressure of the inlet and outlet were measured, and oxygen transfer rate was calculated before operating devices and then every 2 hoursat 4.0L/ min for 12 hours. The plasma free hemoglobin and platelet activation were evaluated every 2 hours as well.Results The satu-ration of outlet was stable at over 98% during the study. The plasma free hemoglobin was lower than 20mg/ dL and the platelet activationwas lower than 8% over 12 hours.Conclusions The wearable artificial cardiopulmonary support device with heparin-coating hollow fi-ber and axial pump conducted the reliable performance in vivo.

关 键 词:动物实验 氧合效率 生物相容性 便携 体外膜肺氧合 

分 类 号:R-332[医药卫生] R318.1

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象