Comparison of two types of visual quality analyzer for the measurement of high order aberrations  被引量:9

Comparison of two types of visual quality analyzer for the measurement of high order aberrations

在线阅读下载全文

作  者:Jing Hao Lin Li Fang Tian Hong Zhang 

机构地区:[1]Department of Cataract, Tianjin Medical University Eye Hospital, Tianjin 300384, China [2]North China University of Science and Technology,Tangshan 063000, Hebei Province, China [3]Tangshan Gongren Hospital, Tangshan 063000, Hebei Province, China

出  处:《International Journal of Ophthalmology(English edition)》2016年第2期292-297,共6页国际眼科杂志(英文版)

摘  要:AIM: To compare the difference and agreement of KR- lW and iTrace for measurement of high order aberrations.METHODS: KR-1W and iTrace were respectively used in a group of healthy people (40 volunteers, 32 eyes) to measure the high order aberration (HOA) of corneal, internal and total ocular. The clinical difference and agreement of two instruments were respectively evaluated by paired-samples 1-test and Bland-Altman analysis. RESULTS: The paired-samples t-test showed that the corneal HOA measured by the two instruments had no statistical differences (P〉0.05); but the internal and total ocular HOA had significant statistical differences (P〈 0.05), and the mean results measured by iTrace were higher than that of KR-1W. However, Bland-Altman analysis revealed that the HOA of corneal and internal were all in 95% limits of agreement; and just one point of total ocular HOA was beyond the 95% limits of agreement. CONCLUSION: KR-1W and iTrace were consistent well in the measurement of corneal, internal and total ocular HOA, especially for the cornea.AIM: To compare the difference and agreement of KR- lW and iTrace for measurement of high order aberrations.METHODS: KR-1W and iTrace were respectively used in a group of healthy people (40 volunteers, 32 eyes) to measure the high order aberration (HOA) of corneal, internal and total ocular. The clinical difference and agreement of two instruments were respectively evaluated by paired-samples 1-test and Bland-Altman analysis. RESULTS: The paired-samples t-test showed that the corneal HOA measured by the two instruments had no statistical differences (P〉0.05); but the internal and total ocular HOA had significant statistical differences (P〈 0.05), and the mean results measured by iTrace were higher than that of KR-1W. However, Bland-Altman analysis revealed that the HOA of corneal and internal were all in 95% limits of agreement; and just one point of total ocular HOA was beyond the 95% limits of agreement. CONCLUSION: KR-1W and iTrace were consistent well in the measurement of corneal, internal and total ocular HOA, especially for the cornea.

关 键 词:KR-IW iTrace high order aberration difference AGREEMENT 

分 类 号:R770.4[医药卫生—眼科]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象