Implantation of trabecular micro-bypass stent using a novel “landing strip” technique  被引量:2

Implantation of trabecular micro-bypass stent using a novel “landing strip” technique

在线阅读下载全文

作  者:Cindy X.Zheng Marlene R.Moster Priyanka Gogte Yang Dai Remy S.Manzi Michael Waisbourd 

机构地区:[1]Glaucoma Research Center, Wills Eye Hospital, Philadelphia,PA 19107, USA

出  处:《International Journal of Ophthalmology(English edition)》2017年第5期738-741,共4页国际眼科杂志(英文版)

摘  要:AIM:To describe a novel technique of creating a landing strip within the trabecular meshwork to guide trabecular micro-bypass stent(iStent) implantation in patients who underwent phacoemulsification.METHODS:Thirty-four eyes from 30 patients who underwent i Stent implantation after phacoemulsification from May 2014 to February 2015 were included in our retrospective study. All iStents were implanted via the "landing strip" technique. A 25-gauge microvitreoretinal blade was used to bisect the trabecular meshwork to less than 1 clock-hour, effectively creating a landing strip. The iStent applicator was pressed along the landing strip and then the stent was released into the trabecular meshwork. RESULTS:Of the 34 eyes with iStent implantation, 27(79.4%) eyes had primary open-angle glaucoma, 6(17.6%) eyes had pseudoexfoliation glaucoma, and 1(2.9%) eye had ocular hypertension. At 6-month follow-up(n=17), the mean number of hypotensive medications decreased from 2.2±1.2 at baseline to 0.8±1.3(P=0.05) and mean intraocular pressure decreased from 19.7±4.1 mm Hg at baseline to 16.7±2.1 mm Hg(P=0.58). Two eyes(5.9%) required subsequent trabeculectomy. CONCLUSION:The "landing strip" technique appears to be an effective way to assist with iS tent implantation.AIM:To describe a novel technique of creating a landing strip within the trabecular meshwork to guide trabecular micro-bypass stent(iStent) implantation in patients who underwent phacoemulsification.METHODS:Thirty-four eyes from 30 patients who underwent i Stent implantation after phacoemulsification from May 2014 to February 2015 were included in our retrospective study. All iStents were implanted via the "landing strip" technique. A 25-gauge microvitreoretinal blade was used to bisect the trabecular meshwork to less than 1 clock-hour, effectively creating a landing strip. The iStent applicator was pressed along the landing strip and then the stent was released into the trabecular meshwork. RESULTS:Of the 34 eyes with iStent implantation, 27(79.4%) eyes had primary open-angle glaucoma, 6(17.6%) eyes had pseudoexfoliation glaucoma, and 1(2.9%) eye had ocular hypertension. At 6-month follow-up(n=17), the mean number of hypotensive medications decreased from 2.2±1.2 at baseline to 0.8±1.3(P=0.05) and mean intraocular pressure decreased from 19.7±4.1 mm Hg at baseline to 16.7±2.1 mm Hg(P=0.58). Two eyes(5.9%) required subsequent trabeculectomy. CONCLUSION:The "landing strip" technique appears to be an effective way to assist with iS tent implantation.

关 键 词:iStent trabecular micro-bypass stent microinvasive glaucoma surgery glaucoma trabecular meshwork 

分 类 号:R779.6[医药卫生—眼科]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象