Correction of Factitious Hyperkalemia in Hemolysed Specimens from Adult Emergency Department Using the Beckman Coulter Unicell DxC 880i Analyzer Derived Factor  

Correction of Factitious Hyperkalemia in Hemolysed Specimens from Adult Emergency Department Using the Beckman Coulter Unicell DxC 880i Analyzer Derived Factor

在线阅读下载全文

作  者:C. Domingues R. Carneiro M. Dias 

机构地区:[1]Clinical Pathology Department of Tdmega e Sousa Hospital Center, Penafiel 4564-007, Portugal

出  处:《Journal of Pharmacy and Pharmacology》2017年第12期902-906,共5页药剂与药理学(英文版)

摘  要:Hemolysis in ED (emergency department) patients is common due to difficult blood draws. Values of serum potassium (K+) become falsely elevated secondary to release of intracellular contents. Objective: The aim of the study was to establish a correction factor for factitious elevated K+ in samples for de adult ED. Methods: We used samples from 125 adult ED patients, in which the 2nd sample was drawn due to hemolysis of the first tube. Results: Firstly, we derived a correction factor expressing an increase in potassium concentration in 0.21 mmol/L (95% confidence interval, 0.17-0.24 mmol/L with p 〈 0.01) for each hemolysis index increment. Conclusions: A reliable correction factor for factitious hyperkalemia in a clinical relevant range exists.

关 键 词:Potassium HEMOLYSIS correction factor. 

分 类 号:R0[医药卫生]

 