瓦里安高剂量率铱源剂量学参数的蒙特卡罗模拟研究  被引量:8

Monte Carlo simulation study of dosimetric parameters of Varian HDR ^192Ir source

在线阅读下载全文

作  者:吴爱林[1] 吴爱东 朱磊[2] 刘云琴 钱立庭 Wu Ailin;Wu Aidong;Zhu Lei;Liu Yunqin;Qian Liting(Tumor Radiotherapy Room,The First Affiliated Hospital of University of Science and Technology of China,Anhui Provincial Cancer Hospital,Hefei 230031,China;School of Physical Sciences,University of Science and Technology of China,Hefei 230026,China;Corresponding author:Email:aidongwu@ 21cn.com)

机构地区:[1]中国科学技术大学附属第一医院安徽省肿瘤医院肿瘤放射治疗室,合肥230031 [2]中国科学技术大学物理学院,合肥230026

出  处:《中华放射医学与防护杂志》2018年第11期859-864,共6页Chinese Journal of Radiological Medicine and Protection

基  金:国家自然科学基金项目(11805198)

摘  要:目的采用美国医学物理师学会(AAPM)和欧洲放射治疗和肿瘤学会(ESTRO)推荐的蒙特卡罗方法对瓦里安GammaMed Plus HDR 192r源的剂量学参数进行模拟研究。方法基于EGSnrc蒙特卡罗软件,建立该型号^192Ir源精确的计算模型。采用公式推导、双线性插值及单位转换等方法,分别得到了单位活度空气比释动能强度、剂量率常数、径向剂量函数以及各向异性函数,并将结果与文献报道数据进行分析比较。结果研究得到的单位活度空气比释动能强度为9.781×10^-8U/Bq,剂量率参数为1.113 cGy·h-^1·U^-1,与文献报道的相差在0.4%以内。本研究的径向剂量函数、各向异性函数与文献数据能较好吻合。结论基于EGSnrc蒙特卡罗软件能对192Ir源剂量学特性进行定量研究,这将为进一步研究后装剂量分布,精确评价临床放疗剂量提供理论依据。Objective To study the dosimetric parameters of Varian GammaMed Plus HDR 192Ir source via Monte Carlo (MC) method based on the recommendation of the American Association of Physicist in Medicine (AAPM) and European Society for Radiotherapy and Oncology(ESTRO). Methods Using the Monte Carlo program EGSnrc, an accurate model of 192Ir source for MC calculations was establish firstly. Through formula derivation, bilinear interpolation and unit conversion, the air kerma strength per unit source activity, the dose rate constant, radial dose function and anisotropy function was obtained then , and compare the result with those in other published studies. Results The air kerma strength per unit source activity was 9.781 × 10^-8 U/Bq. The dose rate constant was 1.113 cGy · h^-1 · U ^-1,with a discrepancy of less than 0.4% compared with result published in other works. Furthermore, the curves of radial dose function and anisotropy function overall agree with the data shown in the literature. Conclusions The feasibility of performing dosimetrie studies of 192Ir source using the MC software EGSnrc was demonstrated. This work provides a theoretical guidance on analysis of the dose distribution of brachytherapy and on evaluation of the dose accuracy of clinical radiotherapy.

关 键 词:近距离后装治疗 ^192IR源 蒙特卡罗模拟 剂量学参数 

分 类 号:R730.55[医药卫生—肿瘤]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象