作 者:孙成淑[1] 张小兰[1] 黄永志[1] Cheng-Shu Sun;Xiao-Lan Zhang;Yong-Zhi Huang(Department of Ophthalmology,West China Hospital,Sichuan University,Chengdu 610000,Sichuan Province,China)
机构地区:[1]四川大学华西医院眼科,中国四川省成都市610000
出 处:《国际眼科杂志》2018年第1期153-155,共3页International Eye Science
摘 要:目的:比较扫频CASIA SS-1000 OCT与Sirius眼前节分析仪两种仪器测量正常人角膜顶点厚度(CCT)和最薄点角膜厚度(TCT)结果的差异性、相关性及一致性,为临床的应用提供理论依据。方法:前瞻性研究。收集近视患者34例,均由同一熟练操作者对受检者右眼进行测量,先采用Sirius眼前节分析仪,再使用SS-1000 OCT重复测量角膜厚度。配对t检验及Bland-Altman用于评估SS-1000 OCT与Sirius眼前节分析仪的角膜顶点及角膜最薄点的一致性。结果:SS-1000 OCT及Sirius角膜地形图仪测量CCT的角膜顶点均值分别为517.62±25.29、518.47±27.23μm。SS-1000 OCT及Sirius眼前节分析仪CCT的角膜最薄点均值分别为513.53±25.06、515.32±26.69μm。两者角膜顶点角膜厚度比较,差异无统计学意义(P>0.05),但两者角膜最薄点的角膜厚度比较,差异具有统计学意义(P<0.05)。两种设备测量的角膜厚度有较好的相关性(r=0.969、0.965,均P<0.05)。做Bland-Altman分析,角膜顶点厚度的95%一致性界限为(-14.22μm,12.52μm),角膜最薄点厚度的95%一致性界限为(-15.61μm,12.03μm),4%(3/68)的点在95%一致性的界限外,但角膜最薄点的一致性差异更大。结论:SS-1000 OCT测量CCT与Sirius眼前节分析仪具有高度的一致性,在临床工作中,可以考虑相互替代。AIM:To compare CASIA SS-1000and Sirius OCT sweep anterior segment analyzer instrument for measuring normal corneal vertex thickness(CCT)and the thinnest corneal thickness(TCT)results the difference,correlation and consistency,and provide a theoretical basis for clinical application.METHODS:This was a prospective study.A total of34normal subjects were collected.The subjects were measured by the same skilled operator.The SS-1000OCT was first used,and then the corneal thickness was measured repeatedly by Sirius anterior segment analyzer.Paired f test and Bland-Altman were used to evaluate the consistency of corneal apex and corneal thinnest point between SS-1000OCT and Sirius anterior segment analyzer.RESULTS:The mean corneal apex measured by SS-1000OCT and Sirius corneal topography were517.62±25.29m m and518.47±27.23m m CCT,respectively.The thinnest points of SS-1000OCT and Sirius anterior segment analyzer GCH'were513.53±25.06m m and515.32±26.69m m,respectively.Paired t test showed that the difference on corneal thickness of vertex was not statistically significant(P>0.05),but the thinnest corneal thickness was statistically significant(P<0.05).Pearson analysis of the two devices,the correlation is0.969,0.965.The results of95%consistency limiting analysis on the corneal vertex thickness by Bland-Altman was(-14.22mm,12.52pm),that of the thinnest corneal thickness was(-15.61nm,12.03Mm),4%(3/68)was out of the95%consistency limiting,but the thinnest corneal thickness was of a little larger differences.CONCLUSION:SS-1000OCT measurement of CCT and Sirius anterior segment analyzer is highly consistent,in clinical work can be considered alternative,but the thinnest point of the cornea can not be replaced each other.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.19.244.116