国产医用加速器XHA1400的X射线剂量和机械性能测试  被引量:8

X-Ray Dose and Mechanical Properties Test of Domestic Medical Accelerator XHA1400

在线阅读下载全文

作  者:吴青南 王运来[2] 鞠忠建[2] 袁树海 戴相昆[2] 宋孟孟 宋迎新 王云 陈富宝 高源 WU Qingnan;WANG Yunlai;JU Zhongjian;YUAN Shuhai;DAI Xiangkun;SONG Mengmeng;SONG Yingxin;WANG Yun;CHEN Fubao;GAO Yuan(School of Physicis Science and Technology,Wuhan University,Wuhan Hubei 430072,China;Department of Radiation Oncology,People’s Liberation Army General Hospital,Beijing 100089,China;Shandong Xinhua Medical Devices Incorporated Company,Zibo Shandong 255000,China)

机构地区:[1]武汉大学物理科学与技术学院,湖北武汉430072 [2]中国人民解放军总医院放疗科,北京100089 [3]山东新华医疗器械股份有限公司,山东淄博255000

出  处:《中国医疗设备》2018年第7期22-26,共5页China Medical Devices

基  金:国家重点研发计划(2016YFC0105715)

摘  要:目的对国产中能医用加速器XHA1400进行X射线剂量输出、机械性能和安全方面的检验,以评估其对三维适形放疗和调强放疗的临床可行性和可靠性。方法以GB15213-2016《医用电子加速器性能和检验方法》和AAPM TG142为基础,结合Elekta等进口设备厂商的相关加速器质量控制标准制定检测项目、检测方法,对XHA1400进行全面的检定。结果除线性要求外,XHA1400各项指标均达到质控标准。结论 XHA1400具有良好的剂量输出准确性和机械性能精度,主要检测指标不输于进口加速器,符合临床进行三维适形放疗和调强放疗的大部分要求。但稍有不足,其剂量输出线性仍需进一步校准以便更好地满足临床调强放疗中出现的小子野剂量输出的需要,以及装备成像质量更高的CBCT系统。Objective In order to evaluate the clinical feasibility and reliability of three-dimensional conformal radiotherapy and intensity modulated radiation therapy of domestic middle-energy medical accelerator XHA1400,we examined the X-ray dose output,mechanical property and safety of it.Methods Based on GB15213-2016 Medical Accelerator Performance and Test Method and AAPM TG142,and referring to the related accelerator quality control standards of imported equipment manufacturers such as Elekta,we conducted a comprehensive verification of the accelerator XHA1400.Results In addition to the results of the linear requirements,XHA1400 indicators had reached quality control standards.Conclusion XHA1400 has good dose output accuracy and mechanical performance accuracy,and the main indicators are not inferior to the imported accelerator.It meets the most of clinic requirement of the three-dimensional conformal radiotherapy and intensity modulated radiotherapy.However,there is still some shortages.Its dose output linearity should be further calibrated in order to better meet the requirement of small field dose output in clinical intensive radiation therapy.What’s more,we should equip the accelerator with cone-beam CT for higher imaging quality.

关 键 词:国产放疗设备 验收测试 剂量特性 机械性能 质量控制 

分 类 号:R730.8[医药卫生—肿瘤] R812[医药卫生—临床医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象