Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study  被引量:4

Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study

在线阅读下载全文

作  者:Güray Altun Gürsel Saka Teyfik Demir Fatma Kübra Erbay Elibol Mehmet Or?un Polat 

机构地区:[1]Department of Orthopaedics and Traumatology, University of Health Sciences Umraniye Training and Research Hospital [2]Department of Orthopaedics and Traumatology, Intercontinental Hisar Hospital [3]Department of Mechanical Engineering, TOBB University of Economics and Technology [4]Department of Industrial Engineering, Kadir Has University, Department of Industrial Engineering

出  处:《World Journal of Orthopedics》2019年第5期219-227,共9页世界骨科杂志(英文版)

摘  要:BACKGROUND The purpose of open reduction and internal fixation of acetabulum posterior wall fractures is to restore anatomical structure and stability of the hip joint, in order to start weight bearing as soon as possible and prevent hip arthrosis; restoration of the anatomy should preserve function of the joint as well. Although "special shaped precontoured plates" have been developed in recent years for surgical treatment of this region, studies comparing the traditional plates with the newly designed precontoured plates are lacking.AIM To evaluate the biomechanical properties of precontoured anatomic buttress and conventional curved reconstruction plates(CCRPs) for posterior wall acetabulum fracture treatment.METHODS Twelve pelvis models were created for testing plate treatment of fracture in the posterior wall of the acetabulum. These 12 pelvis models were used to create 24 hemipelvis models(experimental) by cutting from the sagittal plane and passing over the center of gravity, after which the posterior wall acetabular fractures(of similar type and size) were created. In these experimental models, the right acetabulum was fixed with a 5-hole CCRP, while the left was fixed with a precontoured anatomic buttress plate(PABP). Samples were placed through thetest device and were subjected to static load testing, with a constant testing velocity of 2 mm/min until the load reached 2.3 kN or the acetabular fixation failed. Dynamic tests were also performed with sinusoidal wave load, with a maximal load of 2.3 kN and a load ratio of 0.1.RESULTS The average stiffness values were 460.83 ± 95.47 N/mm for the PABP and 291.99± 118.58 N/mm for the 5-hole CCRP. The precontoured anatomic acetabulum buttress plates had significantly higher rigidity than the CCRPs(P = 0.022). There was a statistically significant difference between the unloaded and 2.3 kN-loaded values of AL(posterosuperior fracture line vertical to the ground surface) and CL(posteroinferior fracture line vertical to the ground surface) parameters for both thBACKGROUND The purpose of open reduction and internal fixation of acetabulum posterior wall fractures is to restore anatomical structure and stability of the hip joint, in order to start weight bearing as soon as possible and prevent hip arthrosis; restoration of the anatomy should preserve function of the joint as well. Although "special shaped precontoured plates" have been developed in recent years for surgical treatment of this region, studies comparing the traditional plates with the newly designed precontoured plates are lacking.AIM To evaluate the biomechanical properties of precontoured anatomic buttress and conventional curved reconstruction plates(CCRPs) for posterior wall acetabulum fracture treatment.METHODS Twelve pelvis models were created for testing plate treatment of fracture in the posterior wall of the acetabulum. These 12 pelvis models were used to create 24 hemipelvis models(experimental) by cutting from the sagittal plane and passing over the center of gravity, after which the posterior wall acetabular fractures(of similar type and size) were created. In these experimental models, the right acetabulum was fixed with a 5-hole CCRP, while the left was fixed with a precontoured anatomic buttress plate(PABP). Samples were placed through thetest device and were subjected to static load testing, with a constant testing velocity of 2 mm/min until the load reached 2.3 kN or the acetabular fixation failed. Dynamic tests were also performed with sinusoidal wave load, with a maximal load of 2.3 kN and a load ratio of 0.1.RESULTS The average stiffness values were 460.83 ± 95.47 N/mm for the PABP and 291.99± 118.58 N/mm for the 5-hole CCRP. The precontoured anatomic acetabulum buttress plates had significantly higher rigidity than the CCRPs(P = 0.022). There was a statistically significant difference between the unloaded and 2.3 kN-loaded values of AL(posterosuperior fracture line vertical to the ground surface) and CL(posteroinferior fracture line vertical to the ground surface) parameters for both th

关 键 词:ACETABULAR FRACTURE Anatomical PLATE Precontoured PLATE POSTERIOR wall FRACTURE BIOMECHANICAL study 

分 类 号:R[医药卫生]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象