基于Nakagami统计参量的活体微波热消融超声监控成像研究  被引量:4

An in vivo study of ultrasonic monitoring imaging of microwave ablation based on Nakagami statistic parameter

在线阅读下载全文

作  者:吴珊 尚少强 王雪维 万明习[1] 张思远 WU Shan;SHANG Shaoqiang;WANG Xuewei;WAN Mingxi;ZHANG Siyuan(Department of Biomedical Engineering,The Key Laboratory of Biomedical InformationEngineering of Ministry of Education,School of Life Science and Technology,Xi’an Jiaotong University,Xi’an 710049,P.R.China)

机构地区:[1]西安交通大学生命科学与技术学院生物医学工程系生物医学信息工程教育部重点实验室

出  处:《生物医学工程学杂志》2019年第3期371-378,共8页Journal of Biomedical Engineering

基  金:国家重点研发计划(2016YFC0104700);国家自然科学基金(61671363,11874297)

摘  要:为了验证采用超声Nakagami统计参量成像对微波热消融治疗过程进行监控成像的可行性,本文提出并研究了采用超声Nakagami统计参量成像对活体猪肝脏和肾脏的微波热消融过程进行监控成像的方法。本文对5头实验猪进行微波热消融治疗,治疗中同步采集超声射频(RF)数据,实验后计算了热凝固区域超声Nakagami统计参量的动态变化,同步构建超声B模式图像和Nakagami统计参量图像,并计算热凝固区域和周边正常组织之间的对比噪声比值(CNR)。微波热消融后,超声Nakagami统计参量图像中出现高亮强回声区域,并随着治疗时间的增加而逐步增大。经计算,肝、肾的超声Nakagami统计参量平均值分别从治疗前的0.78、0.79增大到0.91、0.92;肝、肾的超声Nakagami统计参量图像的CNR平均值从0.49、0.51增加到1.13、0.85,均高于相应超声B模式图像的CNR值。本研究表明,使用超声Nakagami统计参量成像在微波热消融治疗中进行监控具有一定的可行性。This paper explored the feasibility of using ultrasonic Nakagami statistic parameter imaging to evaluate the thermal lesion induced by microwave ablation(MWA)in porcine models.In this paper,thermal lesions were induced in livers and kidneys in 5 swines using a clinical MWA system.During this treatment progress,ultrasonic radiofrequency(RF)data were collected.The dynamic changes of Nakagami parameter in the thermal lesion were calculated,and the ultrasonic B-mode images and Nakagami images were reconstructed simultaneously.The contrast-to-noise ratio(CNR)between the thermal lesion and the surrounding normal tissue was calculated over the MWA procedure.After MWA,a bright hyperechoic region appeared in the ultrasonic Nakagami image as an indicator of the thermal lesion and this bright spot enlarged with lesion development during MWA exposure.The mean value of Nakagami parameter in the liver and kidney increased from 0.78 and 0.79 before treatment to 0.91 and 0.92 after treatment,respectively.During MWA exposure,the mean values of CNR calculated from the Nakagami parameter increased from 0.49 to 1.13 in the porcine liver and increased from 0.51 to 0.85 in the kidney,which were both higher than those calculated from the B-mode images.This in vivo study on porcine models suggested that the ultrasonic Nakagami imaging may provide an alternative modality for monitoring MWA treatment.

关 键 词:微波热消融 超声成像 监控 NAKAGAMI 

分 类 号:R730.5[医药卫生—肿瘤]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象