磁场对Bebig型60Co HDR近距离治疗源剂量分布影响研究  

The impact of a magnetic field on the dose distribution using the Bebig60Co HDR sources

在线阅读下载全文

作  者:崔振国 陈嘉一 云惟康[1] 刘奇 白彦灵[1] Cui Zhenguo;Chen Jiayi;Yun Weikang;Liu Qi;Bai Yanling(Department of Radiotherapy,HaebinMedical University,Harbin 150000,China)

机构地区:[1]哈尔滨医科大学附属肿瘤医院放射物理科,150000

出  处:《中华放射肿瘤学杂志》2020年第3期193-196,共4页Chinese Journal of Radiation Oncology

摘  要:目的计算MRI引导近距离治疗中磁场对Bebig型60Co HDR放射源周围剂量以及碰撞比释动能分布的扰动,判断在不同强度磁场下应用此型号放射源进行近距离治疗可行性。方法基于蒙特卡洛Geant4软件,建立Bebig型60Co近距离治疗源模型,然后置其于尺寸为10cm×10cm×10cm的均匀水模体中,并沿着x、y、z轴向划分大小为0.2mm×0.5mm×0.5mm的体元。在x轴向上分别施加磁场强度为0T、1.5T、3.0T均匀磁场,分别计算x轴向上距离治疗源中心10.0mm范围内碰撞比释动能与剂量分布以及两者比值随距离放射源中心远近变化情况。结果1.5T的磁场对60Co HDR治疗源附近空间范围内剂量分布无影响,而3.0T的磁场导致距离放射源中心r<6mm范围内剂量明显偏高,r=5.4mm处剂量增加了接近40%。碰撞比释动能与剂量比值在1.2mm<r<6.0mm范围内都明显<1,说明3.0T磁场在距离治疗源表面附近更大范围内导致了带电粒子非平衡状态。结论对于Bebig型60Co HDR近距离治疗源,在治疗过程中应用1.5T外部磁场是安全可靠的,而3.0T高场强磁场带来了剂量风险,在投入临床使用前必须进行安全验证评估。Objective To evaluate the impact of an external magnetic field on the dose distribution and electronic disequilibrium region around a Bebig type 60Co HDR brachytherapy source and to judge the feasibility of applying MRI scanner during brachytherapy.Methods The source model was established based on the Monte Carlo package Geant4 software.The simulated geometries consisted of the 60Co source inside a water phantom of 10.0cm×10.0cm×10.0cm in size.The magnetic field strength of the 0T,1.5T and 3.0T was considered,respectively.The voxels with a size of 0.2 mm,0.5 mm and 0.5 mm were established along the x-,y-and z-axis.The influence of the magnetic field on the Kerma(kinetic energy released to matter)distribution and dose distribution within the 10.0mm region from the source center was evaluated.Furthermore,the ratio of the Kerma to dose as a function of the distance to the center source was acquired.Results The 1.5T magnetic field exerted no effect on the dose distribution adjacent to 60Co HDR brachytherapy source,whereas3.0T magnetic field caused significant increase in the dose distribution within r<6 mm from the source center.The dose distribution was increased by 40%at r=5.4 mm from the source center.The ratio of Kerma to dose was less than 1 within the region of 1.2 mm<r<6.0 mm,suggesting that 3.0T magnetic field can lead to electronic disequilibrium within a larger region from the source center.Conclusions For Bebig 60Co HDR brachytherapy source,it is safe and reliable to apply1.5T external magnetic field.Nevertheless,3.0T magnetic field can cause high dose risk.Consequently,safety assessment and verification should be delivered prior to clinical application.

关 键 词:磁场 近距离治疗 剂量 比释动能 

分 类 号:R730[医药卫生—肿瘤]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象