检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:周怀胜[1] 梁婉玲 郎海波 冼志林 晏世刚[1] 孔祥斌[1] 沈沛阳 ZHOU Huaisheng;LIANG Wanling;LANG Haibo;XIAN Zhilin;YAN Shigang;Kong Xiangbin;SHEN Peiyang(Department of Ophthalmology,Foshan Second Pepole̓s Hospital,Foshan 528000,Guangdong,China)
机构地区:[1]佛山市第二人民医院眼科,广东佛山528000
出 处:《山东大学耳鼻喉眼学报》2021年第3期96-99,共4页Journal of Otolaryngology and Ophthalmology of Shandong University
基 金:广东省佛山市医学类科技攻关项目(1920001000762)。
摘 要:目的对比分析欧堡Daytona 200度超广角激光扫描检眼镜和三面镜在孔源性视网膜脱离术前的结果,评估两种检查方法的一致性并分析相关影响因素。方法回顾性分析行玻璃体切除术的孔源性视网膜脱离患者共111例(111眼),术前均进行自然瞳孔下欧堡Daytona 200度超广角激光扫描眼底检查和散瞳后三面镜检查,记录术中和术前检查结果并进行比较分析。结果两种检查方法裂孔检出率的Kappa值为0.447,一致性中等。两种检查方法检出不同类型的视网膜裂孔的数量存在差异。其中差异最大的是圆孔,其他类型裂孔的遗漏相对较少。结论欧堡200度超广角激光扫描检眼镜在孔源性视网膜脱离的术前检查效果与三面镜检查一致性中等,可能会遗漏较周边的视网膜裂孔和黄斑裂孔,尤其是周边视网膜的圆孔。Objective To compare the usefulness of the ultra-widefield laser ophthalmoscope(Daytona)and the three-mirror contact lens for detecting preoperative retinal breaks in a patient with rhegmatogenous retinal detachment,evaluate their consistency,and analyze their related factors.Methods One hundred and eleven eyes of patients with rhegmatogenous retinal detachment were retrospectively reviewed.Retinal images of the patients were obtained with the Optomap ultra-widefield laser ophthalmoscope imaging system and their fundi were examined with a three-mirror contact lens by ophthalmologists.All patients had undergone vitrectomy.The findings from the Optomap retinal images and the three-mirror contact lens examination were compared with the findings of the pars plana vitrectomy(PPV).Results The consistency of the two methods was moderate for detecting rhegmatogenous retinal detachment(Kappa=0.447).The number of different types of retinal holes detected by the two methods was different.The detection of the round hole demonstrated the most significant difference.Few other types of holes were missed for the two methods.Conclusion The ultra-wide-angle laser scanning ophthalmoscope fundus examination system is moderately consistent with the three-mirror contact lens for the preoperative examination of rhegmatogenous retinal detachment,and it may miss more peripheral retinal tears and macular holes,especially peripheral round retinal holes.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:18.219.115.102