作 者:Yoshinori Nakai Kyoko Bessho Yuko Shono Kaori Taoka Yoshihide Nakai
机构地区:[1]Tokai Eye Clinic,399-Hadokoro-cho,Tsu 514-0009,Japan
出 处:《International Journal of Ophthalmology(English edition)》2021年第12期1882-1887,共6页国际眼科杂志(英文版)
摘 要:AIM:To compare the imo perimeter,a new portable head-mounted perimeter unit that enables both eyes to be examined quickly and simultaneously,with the Humphrey field analyzer(HFA)perimeter to investigate correlations and their diagnostic ability in glaucomatous eyes.METHODS:The performance of the equipment in 128 glaucomatous eyes and 40 normal eyes were tested.We investigated the correlations of mean deviation,pattern standard deviation,visual field index,and the sensitivity.RESULTS:Measurements of mean deviation(r=0.886,P<0.001),pattern standard deviation(r=0.814,P<0.001),and visual field index(r=0.871,P<0.001)in both perimeters were strongly and positively correlated.The sensitivities in the imo perimeter were 80.5%for mean deviation,81.2%for pattern standard deviation,and 80.5%in visual field index;those in the HFA were 63.3%for mean deviation,74.5%for pattern standard deviation,and 80.5%for visual field index.Both perimeters demonstrated high diagnostic ability.CONCLUSION:The parameters by the imo and HFA in glaucomatous eyes show strong positive correlations with favorable sensitivity,specificity,and diagnostic ability.However,the difference between imo and HFA results increases with the increase in visual field disturbance.
关 键 词:imo perimeter visual field diagnosis diagnostic validation
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.30