Accuracy of optimized Sirius ray-tracing method in intraocular lens power calculation  被引量:2

在线阅读下载全文

作  者:Li-Qing Wei Ying-Hui Fu Wei-Hua Pan Li Nie Yu Chen Gui-Fang Liu Zhen-Bin Qian 

机构地区:[1]Department of Ophthalmology,the Eye Hospital of Wenzhou Medical University,Hangzhou Xihu Zhijiang Eye Hospital,Hangzhou 310020,Zhejiang Province,China [2]Department of Ophthalmology,the Eye Hospital of Wenzhou Medical University,Wenzhou 325027,Zhejiang Province,China [3]Department of Refractive Surgery,Hangzhou Xihu Zhijiang Eye Hospital,Hangzhou 310024,Zhejiang Province,China

出  处:《International Journal of Ophthalmology(English edition)》2022年第2期228-232,共5页国际眼科杂志(英文版)

基  金:Supported by the Basic Scientific Research Project of Wenzhou (No.2020Y0599)。

摘  要:AIM:To evaluate the accuracy and predictability of ray tracing-assisted intraocular lens(IOL) calculation function in Sirius internal software and further improve the accuracy by optimizing the calculation of predicted lens position(PLP).METHODS:This retrospective study recruited 52 eyes of 49 patients.All of the cases with cataract had undergone phacoemulsification combined with IOL implantation.SRK-T,Haigis formula,and Sirius ray-tracing method were all used for each eye’s IOL calculation.The mean absolute value of prediction error(prediction error=predicted refraction-postoperative refraction) was defined as mean absolute prediction error(MAPE) and was determined for each method.Calculation of PLP was optimized by effective lens position(ELP).Optimized PLP was entered to Sirius internal software again to verify whether the method was improved.RESULTS:Compared with SRK-T and Haigis formulas,less accuracy was shown in Sirius ray-tracing method(P=0.001).The ELP of the IOL moved forward compared to PLP(P<0.001).The MAPE of the ELP-inputted Sirius ray-tracing method was reduced.ELP and PLP were well correlated.Taking ELP as y and PLP given by Sirius soft as x,a linear regression formula y=0.1637 x+3.1741 was concluded(R^(2)=0.1066,P=0.018).It was shown that the optimized Sirius ray-tracing method(optimized PLP entered),compared with SRK-T and Haigis formulas,worked with the same accuracy(P=0.038).CONCLUSION:The original Sirius ray tracing method is not satisfactory enough.However,in normal eyes,the optimized Sirius ray-tracing method in IOL calculation was as accurate as SRK-T and Haigis formulas.

关 键 词:Sirius RAY-TRACING IOL-MASTER CATARACT refractive error predicted IOL position 

分 类 号:R779.6[医药卫生—眼科]

 