Corvis ST与常用眼压计及角膜测厚仪在测量眼压及角膜厚度中的应用  

Observation and Study of Corvis ST,Common Tonometer and Corneal Pachymeter in Measuring Intraocular Pressure and Corneal Thickness

在线阅读下载全文

作  者:黎长金 李雪 陆强[1] Changjin Li;Xue Li;Qiang Lu(Department of Ophthalmology,the second people's Hospital of Foshan,Foshan 528000,China)

机构地区:[1]广东省佛山市第二人民医院眼科,佛山528000

出  处:《中华眼视光学与视觉科学杂志》2022年第4期278-285,共8页Chinese Journal Of Optometry Ophthalmology And Visual Science

摘  要:目的:比较可视化角膜生物力学分析仪(Corvis ST)与三维眼前节分析仪(Pentacam)、非接触式眼压计(NCT)、A型超声角膜测厚仪(A超)、光学生物眼科测量仪(Lenstar)在测量眼压及中央角膜厚度的差异。方法:前瞻性研究。纳入2020年8—12月佛山市第二人民医院屈光术前近视患者110例,将Corvis ST眼压(IOP)及校正眼压(BIOP)、NCT眼压与Pentacam对NCT的校正眼压纳入眼压指标;将Corvis ST的中央角膜厚度、Pentacam的最薄点中央角膜厚度、Lenstar的角膜厚度、A超的角膜厚度纳入角膜厚度指标。眼压与角膜厚度的差异对比采用单因素方差分析法,相关性分析采用Pearson方法,一致性检验采用Bland-Altman。结果:眼压:Corvis ST分别与NCT、Pentacam所测量的眼压值在组间比较差异无统计学意义,其中Corvis ST分别与NCT、Pentacam测量的眼压值比较差异有统计学意义(P=0.019;P=0.03)。IOP与NCT眼压、Pentacam校正眼压呈正相关性(r=0.76,P<0.001;r=0.65,P<0.001),BIOP与NCT眼压、Pentacam校正眼压呈正相关性(r=0.66,P<0.001;r=0.69,P<0.001),IOP与BIOP呈高度正相关性(r=0.92,P<0.001)。IOP与NCT眼压、Pentacam校正眼压、BIOP的95%一致性界限与最大差值绝对值分别为(-2.8~4.3 mmHg,3.8 mmHg)、(-3.1~4.5 mmHg,4.1 mmHg)、(-1.46~1.97 mmHg,1.9 mmHg),BIOP与NCT眼压、Pentacam校正眼压的95%一致性界限与最大差值绝对值分别为(-3.6~4.6 mmHg,4.2 mmHg)、(-3.0~3.9 mmHg,3.4 mmHg)。中央角膜厚度:Corvis ST与A超、Lenstar、Pentacam所测量的中央角膜厚度值在组间比较差异有统计学意义(F=2.67,P=0.046),其中Corvis ST与A超的差异有统计学意义(P=0.017)。Corvis ST与A超、Lenstar、Pentacam所测量的中央角膜厚度值存在着高度正相关性(r=0.96,P<0.001;r=0.98,P<0.001;r=0.98,P<0.001)。Corvis ST与A超、Lenstar、Pentacam对中央角膜厚度的测量值的95%一致性界限与最大差值绝对值分别为(-5.9~24.3μm,23μm)、(-6.4~18.7μm,18μm)、(-10.7~12μm,11μm)�Objective:To compare the difference,correlation and consistency of intraocular pressure and central corneal thickness measured by visual corneal biomechanical analyzer(Corvis ST),three-dimensional anterior segment analyzer(Pentacam),non-contact tonometer(NCT),A ultrasound pachymetry(A-ultrasound)and optical biological ophthalmic measuring instrument(Lenstar).Methods:In this prospective study,110 patients with myopia before refractive surgery in Foshan Second People's Hospital from August to December 2020 were included.Corvis ST intraocular pressure(IOP)and Biomechanical intraocular pressure(BIOP),NCT intraocular pressure and Pentacam corrected intraocular pressure for NCT were included in the intraocular pressure index;the central corneal thickness of Corvis ST,the central corneal thickness of the thinnest point of Pentacam,corneal thickness of Lenstar and corneal thickness of A-ultrasound were included in the corneal thickness index.The difference between intraocular pressure and corneal thickness were compared by One-way ANOVA,Pearson method was used for correlation analysis,and Bland-Altman was used for consistency test.Results:IOP:There was no significant difference in intraocular pressure measured by Corvis ST,NCT and Pentacam between the groups,among them,the difference of intraocular pressure measured by Corvis ST,NCT and Pentacam was statistically significant(P=0.019;P=0.03).IOP was positively correlated with intraocular pressure of NCT and corrected intraocular pressure of Pentacam(r=0.76,P<0.001;r=0.65,P<0.001),BIOP was positively correlated with intraocular pressure of NCT and corrected intraocular pressure of Pentacam(r=0.66,P<0.001;r=0.69,P<0.001),IOP was highly positively correlated with BIOP(r=0.92,P<0.001);the absolute value of 95%consistency limit and maximum difference between IOP and NCT intraocular pressure,Pentacam corrected intraocular pressure and BIOP were(-2.8-4.3 mmHg,3.8 mmHg),(-3.1-4.5 mmHg,4.1 mmHg),(-1.46-1.97 mmHg,1.9 mmHg),the absolute value of 95%consistency limit and maximum diff

关 键 词:可视化角膜生物力学分析仪 三维眼前节分析仪 眼压 中央角膜厚度 

分 类 号:R778.11[医药卫生—眼科]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象