作 者:陈瑞[1] 郑荣[2] 焦姣[1] 燕文敬 彭天浩[1] 邓新立[1] 龚美亮[1] CHEN Rui;ZHENG Rong;JIAO Jiao;YAN Wenjing;PENG Tianhao;DENG Xinli;GONG Meiliang(Department of Clinical Laboratory,Second Medical Center of Chinese PLA General Hospital,Beijing 100853,China;Department of Clinical Laboratory,First Medical Center of Chinese PLA General Hospital,Beijing 100853,China)
机构地区:[1]解放军总医院第二医学中心检验科,北京100853 [2]解放军总医院第一医学中心检验科,北京100853
出 处:《标记免疫分析与临床》2022年第6期1027-1031,共5页Labeled Immunoassays and Clinical Medicine
基 金:军委后勤保障部卫生局保健专项课题(编号:17BJZ45)。
摘 要:目的对胶乳免疫比浊法(比对方法)检测总前列腺特异性抗原(total prostate specific antigen,tPSA)做性能验证,比较其与电化学免疫发光法(参比方法)检测结果的一致性。方法依据CNAS-GL037文件对比对方法检测PSA进行准确度、精密度、线性区间的验证;收集2021年5月至8月解放军总医院第二医学中心剩余血清样本225例,分别由比对方法及参比方法检测总PSA,并对检测结果进行一致性评价。用Passing-Bablok回归、一致性相关系数(CCC)、Bland-Altman图分析2种检测方法的一致性和偏差。结果比对方法检测国家标准物质4个浓度偏倚均在±5%以内,低于厂家声明的15%;重复性和实验室内精密度(CV)不超过2%;临床高低值样本验证1.20~98.12μg/L区间符合临床线性要求。比对方法与参比方法的相关性(r=0.994)、一致性(CCC=0.989)均较好;两种方法检测总PSA平均偏差为-0.617μg/L(-4.36%)。结论比对方法检测总PSA正确度、精密度、线性区间验证均符合要求,比对方法与参比方法检测总PSA具有较好的相关性和一致性。Objective To study the performance of Latex turbidimetric immunoassay(LTIA,comparison method)detecting total prostate specific antigen(PSA),and to evaluate the measurement agreement of LIT with Electro-Chemiluminescence Immunoassay(ECLI,reference method).Methods LIT measurement’s accuracy,precision and linear interval were obtained according to CNAS-GL037.225 residual serum samples were collected from May to August,2021 and total PSA was detected by these two methods.Passing-Bablok regression,concordance correlation coefficient(CCC)and Bland Altman plots were used to analyze the consistency and bias.Results The bias of LIT measuring national reference materials was within±5%,lower than 15%declared by the manufacturer;The repeatability and in laboratory precision(CV)were lower than 2%,and a linear interval between 1.20 and 98.12μg/L passed the validation.The studied method and the reference method showed a good correlation(r=0.994),consistency(CCC=0.989)with an average deviation of-0.617μg/L(-4.36%).Conclusion The studied method shows an accuracy,precision and linear interval which can meet the requirements and also shows a good correlation and consistency with the reference method.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.158