视动性眼球震颤法在儿童视力检查中的应用分析  被引量:1

Analysis of the application of ophthalmic nystagmus in children’s vision examination

在线阅读下载全文

作  者:李长菲 刘桂香[2] 黄启坤 张彩虹 马竟锋[4] 车诚[4] LI Changfei;LIU Guixiang;HUANG Qikun;ZHANG Caihong;MA Jingfeng;CHE Cheng(Department of Ophthalmology,Qilu Hospital(Qingdao),Shandong University,Qingdao 266035,China;Department of Ophthalmology,the Affiliated Hospital of Qingdao University,Qingdao 266003,China;Department of Pediatrics,Qilu Hospital(Qingdao),Shandong University,Qingdao 266035,China;University of Petroleum(East China),Qingdao 266580,China)

机构地区:[1]山东大学齐鲁医院(青岛)眼科,青岛266035 [2]青岛大学附属医院眼科,青岛266003 [3]山东大学齐鲁医院(青岛)儿科,青岛266035 [4]中国石油大学(华东),青岛266580

出  处:《中国眼耳鼻喉科杂志》2022年第5期469-473,共5页Chinese Journal of Ophthalmology and Otorhinolaryngology

摘  要:目的探讨视动性眼球震颤(OKN)法在儿童视力检查中的可行性、准确性与稳定性。方法回顾性对照研究。分别采用OKN法及标准对数视力表对256例6~13岁儿童进行视力检测及分析:①OKN法的总体可测率及不同年龄、性别的差异性;②2种方法检测结果的相关性和一致性;③OKN法检测视力的可重复性。结果不同年龄儿童的可测率为58.00%~87.50%,总体可测率为74.41%±9.16%;不同年龄、性别儿童间可测率差异无统计学意义;OKN法及标准对数视力表总体呈正相关,相关系数r值为0.871(P<0.01);OKN法与标准对数视力表检测结果总体差值为-0.11±0.14,2种方法检测结果的95%一致性界限为-0.38~0.16;OKN法2次检测结果总体差值为-0.02±0.25,2次检测结果的95%一致性界限为-0.51~0.47。结论OKN法用于儿童视力检测具有一定的可行性、准确性以及良好的稳定性。Objective To investigate the feasibility, accuracy and stability of optokinetic nystagmus(OKN) in children’s visual acuity examination. Methods Retrospective control study. A total of 256 children aged 6~13 years were tested by both OKN method and standard logarithmic visual acuity chart. The overall passing rate of OKN method, the differences among different ages and genders, the correlation and consistency of the two methods were detected and the repeatability of OKN method were analyzed. Results The measurability of OKN in children at different ages ranged from 58.00% to 87.50%, and the overall measurability was 74.41%±9.16%. There was no significant difference in the measurability among children at different ages and genders. The results of OKN and standard logarithmic visual acuity chart were positively correlated, and the correlation coefficient r value was 0.871(P < 0.01). The overall difference between OKN method and standard logarithmic visual acuity chart was-0.11±0.14, and the 95% consistency between the two methods was-0.38~0.16. The overall difference between two measurements by OKN method was-0.02±0.25, and the 95% consistency limit was-0.51~0.47. Conclusions The OKN for visual acuity test in children has certain feasibility, accuracy, and good stability.

关 键 词:视动性眼球震颤 视力 准确性 儿童视力 稳定性 

分 类 号:R777.44[医药卫生—眼科]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象