作 者:张慧 张惠珠 牛红霞 张翠霞[1] 祝艳蕊 ZHANG Hui;ZHANG Huizhu;NIU Hongxia;ZHANG Cuixia;ZHU Yanrui(Heze Center for Disease Control and Prevention,Heze 274000,China;Heze Testing Institute for Food and Drug Control,Heze 274000,China)
机构地区:[1]菏泽市疾病预防控制中心,山东菏泽274000 [2]菏泽市食品药品检验检测研究院,山东菏泽274000
出 处:《现代食品》2022年第18期150-153,162,共5页Modern Food
摘 要:目的:对比微波消解和快速消解处理方法的效果,探讨两种方法的适用性和可靠性,建立快速、高效测定突发食品重金属中毒、质控考核中铅、镉、砷的方法。方法:采用微波消解和快速消解分别进行样品前处理,ICP-MS法测定,从检出限、准确度、精密度以及加标回收率等方面进行对比分析。结果:两种测定方法的检出限无显著差异,标准物质的测定值均在认定值范围之内,RSD <10%,加标回收率在88.9%~109.8%,盲样中两方法测定铅、镉、总砷的结果差异无统计学意义(Pb t=-2.423,P>0.05;Cd t=1.275,P>0.05;As t=-2.200,P>0.05),反馈的盲样考核结果为满意(Pb 0.150 mg·kg^(-1),Z=0.34;Cd 1.02 mg·kg^(-1),Z=-0.56;As 0.104 mg·kg^(-1),Z=0.57)。结论:快速消解试剂用量少、耗时短、操作简单,在突发应急检测、盲样考核中可代替微波消解进行各元素的测定。Objective: This research makes a comparative study on the effects of microwave digestion and rapid digestion, and probes into the applicability and reliability of the two methods, so as to settle on a rapid and efficient method for determining lead, cadmium and arsenic in emergent food heavy metal poisoning and quality control assessment. Method:Microwave digestion and rapid digestion are adopted to pretreat the samples, and ICP-MS is used to measure, whereby comparative analysis is made in terms of detection limit, accuracy, precision, and recovery rate. Result: Microwave digestion and rapid digestion yield to detection limit with no significant difference, and the measured values of standard substances are all within the range of certified values, RSD<10%, and the recovery of standard addition is between 88.9%~109.8%.In blind samples, the measurements of lead, cadmium and total arsenic in the two methods present no significant differences and no statistical significance(Pb t=-2.423,P>0.05;Cd t=1.275,P>0.05;As t=-2.200,P>0.05) and the feedback blind sample assessment result is satisfactory(Pb 0.150 mg·kg^(-1), Z=0.34;Cd 1.02 mg·kg^(-1), Z=-0.56;As 0.104 mg·kg^(-1), Z=0.57).Conclusion: The rapid digestion features less reagent dosage, shorter time and simpler operation, hence it can replace the microwave digestion in emergent situations and blind sample assessments to determine the elements.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.4