全帧CCD探测器CT扫描数据读出与拖影校正研究  被引量:1

Study on Data Readout and Smear Correction of Full-frame CCD Detector for CT Scanning

在线阅读下载全文

作  者:安康[1,2] 吴石琳 杜宇 段晓礁[1] 李汶芳[1] 周日峰 王珏[1] AN Kang;WU Shilin;DU Yu;DUAN Xiaojiao;LI Wenfang;ZHOU Rifeng;WANG Jue(ICT Research Center,Key Laboratory of Optoelectronic Technology and Systems,Ministry of Education,Chongqing University,Chongqing 400044,China;State Key Laboratory of Mechanical Transmission,Chongqing University,Chongqing 400044,China;Institute of Chemical Materials,China Academy of Engineering Physics,Mianyang 621000,China)

机构地区:[1]重庆大学光电技术及系统教育部重点实验室ICT研究中心,重庆400044 [2]重庆大学机械传动国家重点实验室,重庆400044 [3]中国工程物理研究院化工材料研究所,四川绵阳621000

出  处:《原子能科学技术》2022年第10期2203-2213,共11页Atomic Energy Science and Technology

基  金:国家自然科学基金重大科研仪器研制项目(11827809);重庆大学机械传动国家重点实验室课题(SKLMT-ZZKT-2021M04)。

摘  要:以光纤耦合全帧CCD方式制作的X射线探测器在实验室高精度X射线CT扫描中广泛应用。由于X射线的强穿透特性,全帧CCD探测器通常工作在无快门模式,图像拖影是该模式下必定存在的伪影。本文针对全帧CCD探测器CT扫描图像拖影校正进行研究,提出了非曝光行校正法校正投影图像拖影,避免了利用暗像元校正法校正投影图像拖影时可能会带来的条形伪影;针对动态曝光CT扫描,设计了基于辐射屏蔽的投影图像感兴趣区域快速读出模式,通过降低电荷转移时间有效抑制投影图像拖影的产生,提高探测器有效信号动态范围。实验结果表明:应用非曝光行校正法能有效校正投影图像拖影、提高投影图像质量;采用投影图像感兴趣区域快速读出模式可有效解决动态曝光CT扫描图像拖影校正难题。X-ray detectors made by optical fiber coupled with full-frame CCD are widely used in high-precision X-ray CT scanning of laboratory.Because of the strong penetration of X-rays,the conventional shutter can’t realize ray shielding,so the full-frame CCD detector usually works in the shutter-free mode which can’t avoid generating image smears.Due to image smears,the image observation is seriously affected and the dynamic range of effective data of the CCD detector is reduced.The smears of projection images of full-frame CCD detector must be corrected.In order to correct smears of projection images in CT scanning application,the characteristics of image smear of full-frame CDD in standard readout mode and charge flush readout mode were studied in this paper.And the applicability of the widely used smear correction method using line pixels data of the optical black region to the smear correction of X-ray radiation imaging was studied too.It is found that the application of dark pixels correction method to correct the radiation projection image may bring thin strip artifacts.The results come from two main reasons.Firstly,the fluctuation of pixels data can’t be effectively suppressed by calculating the mean value,because the number of lines of pixels data of the optical black region is too small.Secondly,the fluctuation of line pixels data of the optical black region will increase due to the radiation interference during integration.An improved smear correction method using non-exposed line pixels data was proposed in this paper to avoid the two shortcomings above.Using this method,the image smear is corrected by using the non-exposed line pixels data which aren’t exist during integration,so the radiation interference during integration is well solved.In addition,the stability of correction data is improved by calculating the mean value of more non-exposed line pixels data because the number of non-exposed line pixels data can be customized on demand.Further study of dynamic exposure CT scanning imaging shows tha

关 键 词:全帧CCD 非曝光行 感兴趣区域 行转移拖影 等待读出拖影 

分 类 号:TL81[核科学技术—核技术及应用]

 