机构地区:[1]甘肃省康复中心医院白内障复明中心,甘肃省兰州市730000
出 处:《国际眼科杂志》2023年第2期273-277,共5页International Eye Science
摘 要:目的:探讨年龄相关性白内障患者应用新型扫频源光学相干断层扫描仪(SS-OCT)与Scheimpflug眼前节分析仪测量角膜屈光力、角膜散光的Bland-Altman分析。方法:选取2021-01/12我院收治的年龄相关性白内障患者177例282眼。分别采用SS-OCT与Scheimpflug眼前节分析仪测量角膜前表面、后表面、全角膜的陡峭轴屈光力(Ks)、平坦轴屈光力(Kf)、平均角膜屈光力(Km)、角膜散光、散光轴向。所有参数进行配对样本t检验、组内重复性检验、Pearson相关性和Bland-Altman一致性分析。结果:SS-OCT与Scheimpflug眼前节分析仪测量的角膜前表面屈光力Ks、Kf、Km均无差异(P>0.05)。Scheimpflug眼前节分析仪测量的角膜后表面以及全角膜屈光力Ks、Kf、Km均大于SS-OCT测量值(P<0.05)。SS-OCT与Scheimpflug眼前节分析仪测量的角膜前表面、角膜后表面以及全角膜的散光度数和轴向均无差异(P>0.05)。所有参数的组内相关系数(ICC)均大于0.88,提示组内重复性较好。SS-OCT与Scheimpflug眼前节分析仪测量的角膜前表面、角膜后表面以及全角膜的Ks、Kf、Km、散光度数和轴向均呈正相关(P<0.05)。Bland-Altman一致性分析显示,SS-OCT与Scheimpflug眼前节分析仪测量的角膜前表面、角膜后表面以及全角膜的Ks、Kf、Km、散光度数和轴向的一致性较好(P<0.05)。结论:新型SS-OCT与Scheimpflug眼前节分析仪测量的角膜屈光力、角膜散光参数具有高度一致性,可适用于年龄相关性白内障患者角膜屈光力及角膜散光的诊断。AIM:To explore Bland-Altman analysis of corneal curvature and astigmatism measured by new swept-source optical coherence tomography(SS-OCT)and Scheimpflug anterior segment analyzer in patients with age-related cataract.METHODS:A total of 177 patients(282 eyes)with age-related cataract admitted to the hospital between January and December 2021 were enrolled.The steep-axis curvature(Ks),flat-axis curvature(Kf),mean corneal curvature(Km),corneal astigmatism and astigmatism axis of anterior corneal surface,posterior surface and the whole cornea were measured by SS-OCT and Scheimpflug anterior segment analyzer respectively.All parameters were detected by paired sample t-test,intra-group repeatability test,Pearson correlation and Bland-Altman consistency analysis.RESULTS:There was no significant difference in Ks,Kf and Km of anterior corneal surface measured by SS-OCT and Scheimpflug anterior segment analyzer(P>0.05).Ks,Kf and Km of posterior corneal surface and whole cornea measured by Scheimpflug anterior segment analyzer were all greater than those measured by SS-OCT(P<0.05).There was no significant difference in astigmatism and axial values of anterior corneal surface,posterior surface and whole cornea measured by SS-OCT and Scheimpflug anterior segment analyzer(P>0.05).The intraclass correlation coefficient(ICC)of all parameters was greater than 0.88,indicating a good intra-group repeatability.Ks,Kf,Km,astigmatism and axis of anterior corneal surface,posterior surface and whole cornea measured by SS-OCT were positively correlated with those measured by Scheimpflug anterior segment analyzer(P<0.05).Bland-Altman consistency analysis showed that Ks,Kf,Km,corneal astigmatism and axis of anterior corneal surface,posterior surface and whole cornea measured by SS-OCT were highly consistent with those measured by Scheimpflug anterior segment analyzer(P<0.05).CONCLUSION:The corneal curvature and astigmatism parameters measured by new SS-OCT are highly consistent with those measured by Scheimpflug anterior segment analyzer,
关 键 词:白内障 新型扫频源光学相干断层扫描仪(SS-OCT) 眼前节分析仪 角膜屈光力 散光
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...