作 者:邓和平[1] 张宇 杨学刚[1] 卢予婕 王丹[1] 董晓蕾[1] 何雅坤[1] 陈佳 王闽[1] DENG He-ping;ZHANG Yu;YANG Xue-gang(Department of Radiology,Sichuan cancer Hospital&institute,Chengdu 610000,China)
机构地区:[1]四川省肿瘤医院研究所影像诊断部,成都610000 [2]上海联影医疗科技股份有限公司磁共振临床培训部,上海201800
出 处:《放射学实践》2023年第5期631-635,共5页Radiologic Practice
基 金:北京医学奖励基金会课题(YXJL-2020-0972-0424)。
摘 要:目的:探讨光梭径向采集(uCSR)技术在屏气配合欠佳患者进行肝脏动态增强扫描中的应用价值。方法:搜集35例无法配合屏气检查的患者作为研究对象,在自由呼吸状态下采用uCSR序列进行肝脏动态增强扫描,该序列扫描结束后立即进行3D-VIBE序列扫描。由两位高年资医师采用5分法分别对uCSR重建的各期相准确性、uCSR序列延迟期与3D-VIBE序列延迟期总体图像质量及呼吸运动伪影抑制效果进行评分。由两位高年资技师对uCSR序列延迟期和3D-VIBE序列延迟期图像进行信噪比(SNR)及对比噪声比(CNR)值测量。结果:两位医师对uCSR重建的各期相准确性评分均高于4.0分(分别为4.86±0.355、4.83±0.382),表明uCSR序列能获得准确的肝脏动态增强扫描期相。对3D-VIBE序列和uCSR序列扫描的延迟期图像进行总体质量评分,结果显示uCSR图像明显优于3D-VIBE(Z=-5.241,P=0.000)。对3D-VIBE序列与uCSR序列延迟期图像的呼吸运动伪影抑制效果进行评分,结果显示uCSR图像更优(Z=-5.279,P=0.000)。对两组图像质量进行客观评价,结果显示3D-VIBE图像的SNR稍优于uCSR(Z=-2.359,P=0.018),CNR差异无统计学意义(Z=-1.753,P=0.08)。结论:对于屏气配合欠佳的患者进行肝脏动态增强扫描时,uCSR序列可以获得良好的图像,具有一定临床应用价值。Objective:To investigate the value of United imaging Compressed Sensing with Radial acquisition(uCSR)technology in liver dynamic contrast-enhanced(DCE)scanning in patients with poor breath-holding ability.Methods:35 patients who failed to cooperate with the breath-hold examination were included in this study.They received DCE scanning with uCSR sequence in free-breathing state,followed immediately with 3D-VIBE sequence scanning.Two senior physicians scored the phase accuracy of uCSR reconstruction,the overall quality of the delayed phase images of uCSR sequence and 3D-VIBE sequence,the suppressing effect of respiratory motion artifacts using the 5-point scales.Two senior technicians measured signal-to-noise ratio(SNR)and contrast-to-noise ratio(CNR)of the delay phase images of uCSR sequence and 3D-VIBE sequence.Results:The mean scores of each phase accuracy for uCSR reconstruction rated by the two physicians are significantly higher than 4.0(4.86±0.355 and 4.83±0.382,respectively),indicating that the uCSR sequence can produce accurate phases.According to scores for the overall quality of the delay phase images of uCSR sequence and 3D-VIBE sequence,uCSR images are significantly better than 3D-VIBE(Z=-5.241,P=0.000).According to comparison between the delayed phase images of uCSR sequence and 3D-VIBE sequence in the suppressing effect on respiratory motion artifacts,uCSR images are better(Z=-5.279,P=0.000).For objective evaluation of image quality,3D-VIBE had slightly higher SNR than uCSR group(Z=-2.359,P=0.018),while CNR showed no significant difference(Z=-1.753,P=0.08).Conclusion:uCSR sequence can obtain excellent liver DCE images for patients with poor breath-holding and has potential in clinical application.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222