作 者:Vivian Wing-Man Ho Hong Chau Nguyen Siu-Hong Wong Anita Lai-Wah Li Kenneth Kai Wang Li
机构地区:[1]Department of Ophthalmology,United Christian Hospital,Kwun Tong,Hong Kong SAR,China [2]Department of Ophthalmology,Tseung Kwan O Hospital,Hong Kong SAR,China [3]Department of Ophthalmology,School of Clinical Medicine,LKS Faculty of Medicine,The University of Hong Kong,Hong Kong SAR,China
出 处:《Annals of Eye Science》2023年第3期59-69,共11页眼科学年鉴(英文)
摘 要:Background:To compare two swept-source optical coherence tomography(SS-OCT)biometers,IOLMaster 700 and ANTERION.Methods:This is a retrospective study.Biometric measurements of cataract patients performed between March and July 2021 in the Department of Ophthalmology,United Christian Hospital,Hong Kong,were reviewed.Patients scheduled for cataract surgery were measured with both SS-OCT devices on the same day.The following biometry parameters were compared:keratometry(K),total keratometry(TK),axial length(AL),central corneal thickness(CCT),anterior chamber depth(ACD),lens thickness(LT),white-to-white(WTW)and the predicted intraocular lens(IOL)power to achieve emmetropia.To assess the agreement between the devices,Bland-Altman analysis with 95%limits of agreement(LoA)were used.Results:In total,92 eyes of 47 subjects were measured with both devices.There were statistically significant differences between the two biometers for most measurements(P<0.05)except for flat K,AL and IOL power when using the right eyes for analysis.For the left eyes,there were statistically significant differences in the measurements from the two biometers in all parameters except for flat and steep K.The ANTERION did not obtain ACD,AL and LT in 2(2.17%),1(1.09%)and 5 cases(5.43%)respectively.Conclusions:The two biometers showed a clinically acceptable agreement in most parameters.Comparisons showed significant differences in most parameters but not clinically relevant except for the TK and WTW,and these two parameters should not be used interchangeably between the devices.
关 键 词:Ocular biometry CATARACT swept-source optical coherence ANTERION IOLMaster 700
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.40