EDGE SP1000单孔手术机器人辅助腹腔镜下猪输尿管部分切除联合端端吻合术的可行性研究  被引量:1

Feasibility study of EDGE SP1000 single hole surgical robot assisted laparoscopic partial ureterectomy with end-to-end anastomosis in pigs

在线阅读下载全文

作  者:刘成[1] 赖聪 黄健[1] 王建辰 罗茜芸 许可慰[1] Cheng Liu;Cong Lai;Jian Huang;Jianchen Wang;Xiyun Luo;Kewei Xu(Department of Urology,Sun Yat-sen Memorial Hospital,Sun Yat-sen University,Guangzhou 510120,China;Shenzhen Jingfeng Medical Technology Co.,LTD.,Guangdong 518100,China)

机构地区:[1]中山大学孙逸仙纪念医院泌尿外科,广州510120 [2]深圳市精锋医疗科技股份有限公司,广东518100

出  处:《中华腔镜泌尿外科杂志(电子版)》2023年第6期642-646,共5页Chinese Journal of Endourology(Electronic Edition)

基  金:广东省重点领域研发计划"高端医疗器械"重点专项(2020B111114002)。

摘  要:目的利用活体猪模型评估EDGE SP1000单孔机器人辅助腹腔镜输尿管部分切除联合端端吻合术的安全性和可行性。方法利用EDGE SP1000单孔机器人手术系统开展腹腔镜下猪输尿管部分切除联合端端吻合术。监测手术过程中猪的生命体征,记录手术困难点和并发症,观察术后2周猪的存活情况。结果所有猪均存活至术后两周,手术过程中未出现并发症,也未遇到技术困难。5头猪建立手术通道的平均时间为(15.0±1.2)min,安装EDGE SP1000单孔手术机器人系统时间为(5.6±0.6) min,输尿管部分切除时间为(8.8±1.3) min,输尿管端端吻合时间为(39.2±2.3) min,总手术时间为(89±5) min,术中出血量为(1.20±0.24)mL。使用该手术系统的外科医师和助手均对系统的工作给予了高度评价。结论本研究证实EDGE SP1000单孔机器人手术系统用于猪输尿管部分切除联合端端吻合术是安全可行的。下一步拟开展EDGE SP1000系统在其他外科手术中的安全性和有效性评价,提供更多可靠的临床前数据支撑。ObjectiveTo evaluate the safety and feasibility of using the EDGE SP1000 single-port surgical robot for laparoscopic partial ureterectomy with end-to-end anastomosis in a live pig model.MethodsThe EDGE SP1000 single-port robotic surgical system was used to perform laparoscopic partial ureterectomy with end-to-end anastomosis in pigs. Vital signs of the pigs were monitored during the surgery, and surgical difficulties and complications were recorded. The survival of the pigs was observed for two weeks post-operatively.ResultsAll pigs survived for two weeks after the surgery, with no surgical complications or technical difficulties encountered during the procedure. The average time for establishing the surgical access in five pigs was (15.0±1.2) min, while the installation time for the EDGE SP1000 single-port surgical robot system was also (5.6±0.6) min. The time for ureterectomy was (8.8±1.3) min, the time for end-to-end anastomosis of the ureter was (39.2±2.3) min, and the total surgical time was (89±5) min, and intraoperative blood loss was (1.20±0.24) mL. The surgical team highly praised the performance of the surgical system.ConclusionThis study confirms that the EDGE SP1000 single-port robotic surgical system is safe and feasible for laparoscopic partial ureterectomy with end-to-end anastomosis in pigs. The next step is to evaluate the safety and effectiveness of the EDGE SP1000 system in other surgical procedures, providing more reliable preclinical data support.

关 键 词:单孔手术 机器人 机器人辅助腹腔镜 输尿管部分切除 端端吻合 活体猪 

分 类 号:R699[医药卫生—泌尿科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象